Single-Particle Soot Photometer (SP2) Instrument Handbook
- Brookhaven National Lab. (BNL), Upton, NY (United States)
The SP2 is an instrument that measures, in situ, the time-dependent scattering and incandescence signals produced by individual BC-containing particles as they travel through a continuous-wave laser beam. Any particle traversing the laser beam will scatter light, and the BC component of a BC-containing particle will absorb some of the laser energy until its temperature is raised to the point at which it incandesces (hereafter we adopt the standard terminology of the SP2 community and denote any substance determined by the SP2 to be BC as refractory black carbon (rBC)). The amplitude of the rBC incandescence signal is related to the amount of refractory material contained in the illuminated particle. By binning the individual incandescence signals per unit sample volume, the mass concentration [ng/m3] of rBC can be derived. By binning the individual signals by volume equivalent diameter the size distribution (dN/dlogDVED) per unit time can be derived. The rBC mass loading per unit time and the rBC size distribution unit time are the core data products produced by the SP2. Additionally, the scattering channel can be used to provide information on the rBC particle population-based mixing states within ambient aerosols. However, this data product is produced on a requested-basis since additional detailed analysis and QC/QA must be conducted.
- Research Organization:
- DOE Office of Science Atmospheric Radiation Measurement (ARM) Program (United States)
- Sponsoring Organization:
- USDOE Office of Science (SC), Biological and Environmental Research (BER)
- DOE Contract Number:
- AC05-7601830
- OSTI ID:
- 1344179
- Report Number(s):
- DOE/SC-ARM-TR-169
- Country of Publication:
- United States
- Language:
- English
Similar Records
ACMEV-SP2 (Single Particle Soot Photometer)
Single Particle Soot Photometer (SP2) BBOP