skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Multi-energy x-ray detector calibration for Te and impurity density (nZ) measurements of MCF plasmas

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4960602· OSTI ID:1309457

Here, soft x-ray detection with the new "multi-energy" PILATUS3 detector systems holds promise as a magnetically confined fusion (MCF) plasma diagnostic for ITER and beyond. The measured x-ray brightness can be used to determine impurity concentrations, electron temperatures, n2eZeff products, and to probe the electron energy distribution. However, in order to be effective, these detectors which are really large arrays of detectors with photon energy gating capabilities must be precisely calibrated for each pixel. The energy-dependence of the detector response of the multi-energy PILATUS3 system with 100 K pixels has been measured at Dectris Laboratory. X-rays emitted from a tube under high voltage bombard various elements such that they emit x-ray lines from Zr-Lα to Ag-Kα between 1.8 and 22.16 keV. Each pixel on the PILATUS3 can be set to a minimum energy threshold in the range from 1.6 to 25 keV. This feature allows a single detector to be sensitive to a variety of x-ray energies, so that it is possible to sample the energy distribution of the x-ray continuum and line-emission. PILATUS3 can be configured for 1D or 2D imaging of MCF plasmas with typical spatial energy and temporal resolution of 1 cm, 0.6 keV, and 5 ms, respectively.

Research Organization:
Princeton Plasma Physics Lab. (PPPL), Princeton, NJ (United States)
Sponsoring Organization:
USDOE
Grant/Contract Number:
AC02-09CH11466
OSTI ID:
1309457
Report Number(s):
5266; RSINAK
Journal Information:
Review of Scientific Instruments, Vol. 87, Issue 11; ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)Copyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 7 works
Citation information provided by
Web of Science

References (6)

X-ray imaging crystal spectroscopy for use in plasma transport research journal November 2012
Impurity transport studies in NSTX neutral beam heated H-mode plasmas journal July 2009
Novel energy resolving x-ray pinhole camera on Alcator C-Mod journal October 2012
Characterization and Calibration of PILATUS Detectors journal June 2009
Edge multi-energy soft x-ray diagnostic in Experimental Advanced Superconducting Tokamak journal December 2015
X-ray observations of 2 l - nl ′ transitions from Zr, Nb, Mo, and Pd in near-neonlike charge states journal June 1996

Cited By (2)

Pixel-to-pixel variation on a calibrated PILATUS3-based multi-energy soft x-ray detector journal October 2018
In situ wavelength calibration system for the X-ray Imaging Crystal Spectrometer (XICS) on W7-X journal October 2018