Title: Development of a portable mass spectrometric system for determination of isotopic composition of solid uranium samples using fluorine volatilization. Final report

Technical Report ·
DOI:https://doi.org/10.2172/10116230· OSTI ID:10116230

Using hardware and materials supplied by LANL, a prototype quadrupole mass spectrometer system designed for portable field analysis of isotopic composition of solid uranium samples was assembled and tested. The system contained the capability for fluorine volatilization of solid uranium samples with gas introduction, which was successfully tested and demonstrated using 100 mg samples of U{sub 3}O{sub 8}. Determination of precision and accuracy for measuring isotopic composition was performed using isotopic standards. Use with soil samples containing uranium were also attempted. Silicates in the soil forming SiF{sub 4} were found to be a kinetic bottleneck to the formation of UF{sub 6}. This could be avoided by performing some sort of chemical separation as a pre-treatment step, which was demonstrated using nitric acid.