Fabrication of Gamma Detectors Based on Magnetic Ag:Er Microcalorimeters
- Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
- Univ. of New Mexico, Albuquerque, NM (United States)
- STAR Cryoelectronics, Santa Fe, NM (United States)
This report discusses the photolithographic fabrication of ultra-high resolution gamma-ray detectors based on magnetic microcalorimeters (MMCs). The MMC uses a novel Er-doped silver sensor (Ag:Er) that is expected to have higher sensitivity than the Er-doped gold (Au:Er) sensors currently in use. The MMC also integrates the first-stage SQUID preamplifier on the same chip as the MMC gamma detector to increase its signal-to-noise ratio. In addition, the MMC uses a passive Ta-Nb heat switch to replace one of the common long-term failure points in earlier detectors. This report discusses the fabrication process we have developed to implement the proposed improvements.
- Research Organization:
- Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC52-07NA27344
- OSTI ID:
- 1256452
- Report Number(s):
- LLNL-TR-691117; TRN: US1700210
- Country of Publication:
- United States
- Language:
- English
Similar Records
Fabrication of Gamma Detectors Based on Magnetic Ag:Er Microcalorimeters
Design of a Dry Dilution Refrigerator for MMC Gamma Detector Arrays
Magnetic Microcalorimeter (MMC) Gamma Detectors with Ultra-High Energy Resolution
Technical Report
·
Wed Nov 25 00:00:00 EST 2015
·
OSTI ID:1256452
Design of a Dry Dilution Refrigerator for MMC Gamma Detector Arrays
Technical Report
·
Mon Apr 03 00:00:00 EDT 2017
·
OSTI ID:1256452
Magnetic Microcalorimeter (MMC) Gamma Detectors with Ultra-High Energy Resolution
Technical Report
·
Wed Jan 09 00:00:00 EST 2019
·
OSTI ID:1256452