Experimental Design for a Macrofoam Swab Study Relating the Recovery Efficiency and False Negative Rate to Low Concentrations of Two Bacillus anthracis Surrogates on Four Surface Materials
Abstract
This report describes the experimental design for a laboratory study to quantify the recovery efficiencies and false negative rates of a validated, macrofoam swab sampling method for low concentrations of Bacillus anthracis Sterne (BAS) and Bacillus atrophaeus (BG) spores on four surface materials (stainless steel, glass, vinyl tile, plastic light cover panel). Two analytical methods (plating/counting and polymerase chain reaction) will be used. Only one previous study has investigated false negative as a function of affecting test factors. The surrogates BAS and BG have not been tested together in the same study previously. Hence, this study will provide for completing gaps in the available information on the performance of macrofoam swab sampling at low concentrations.
- Authors:
- Publication Date:
- Research Org.:
- Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
- Sponsoring Org.:
- USDOE
- OSTI Identifier:
- 1130248
- Report Number(s):
- PNNL-23094
400904120
- DOE Contract Number:
- AC05-76RL01830
- Resource Type:
- Technical Report
- Country of Publication:
- United States
- Language:
- English
- Subject:
- Bacillus anthracis; Swab sampling; False negative rate; Recovery Efficiency
Citation Formats
Piepel, Gregory F., and Hutchison, Janine R. Experimental Design for a Macrofoam Swab Study Relating the Recovery Efficiency and False Negative Rate to Low Concentrations of Two Bacillus anthracis Surrogates on Four Surface Materials. United States: N. p., 2014.
Web. doi:10.2172/1130248.
Piepel, Gregory F., & Hutchison, Janine R. Experimental Design for a Macrofoam Swab Study Relating the Recovery Efficiency and False Negative Rate to Low Concentrations of Two Bacillus anthracis Surrogates on Four Surface Materials. United States. https://doi.org/10.2172/1130248
Piepel, Gregory F., and Hutchison, Janine R. 2014.
"Experimental Design for a Macrofoam Swab Study Relating the Recovery Efficiency and False Negative Rate to Low Concentrations of Two Bacillus anthracis Surrogates on Four Surface Materials". United States. https://doi.org/10.2172/1130248. https://www.osti.gov/servlets/purl/1130248.
@article{osti_1130248,
title = {Experimental Design for a Macrofoam Swab Study Relating the Recovery Efficiency and False Negative Rate to Low Concentrations of Two Bacillus anthracis Surrogates on Four Surface Materials},
author = {Piepel, Gregory F. and Hutchison, Janine R.},
abstractNote = {This report describes the experimental design for a laboratory study to quantify the recovery efficiencies and false negative rates of a validated, macrofoam swab sampling method for low concentrations of Bacillus anthracis Sterne (BAS) and Bacillus atrophaeus (BG) spores on four surface materials (stainless steel, glass, vinyl tile, plastic light cover panel). Two analytical methods (plating/counting and polymerase chain reaction) will be used. Only one previous study has investigated false negative as a function of affecting test factors. The surrogates BAS and BG have not been tested together in the same study previously. Hence, this study will provide for completing gaps in the available information on the performance of macrofoam swab sampling at low concentrations.},