Empirical deadtime corrections at synchrotron sources.
Journal Article
·
· J. Synchrotron Radiat.
An experimental comparison of models for performing dead-time corrections of photon-counting detectors at synchrotron sources is presented. The performance of several detectors in the three operating modes of the Advanced Photon Source is systematically compared, with particular emphasis on asymmetric fill patterns. Several simple and well known correction formulas are evaluated. The results demonstrate the critical importance of detector speed and synchrotron fill pattern in selecting the proper dead-time correction.
- Research Organization:
- Argonne National Lab. (ANL), Argonne, IL (United States)
- Sponsoring Organization:
- USDOE Office of Science (SC)
- DOE Contract Number:
- DE-AC02-06CH11357
- OSTI ID:
- 981321
- Report Number(s):
- ANL/XSD/JA-61831; JSYRES; TRN: US1003948
- Journal Information:
- J. Synchrotron Radiat., Vol. 15, Issue 2008; ISSN 0909-0495
- Country of Publication:
- United States
- Language:
- ENGLISH