Title: [sup 3]He neutron detector performance in mixed neutron gamma environments

A test program of the performance of 3He neutron proportional detectors with varying gas pressures, and their response to lligh level gamma-ray exposure in a mixed neutrodgamma environment, ha$ been performed Our intent was to identie the optimal gas pressure to reduce the gamma-ray sensitivity of these detectors. These detectors were manufxtured using materials to minimize their gamma response. Earlier work focused on 3He fill pressures of four atmospheres and above, whereas the present work focuses on a wider range of pressures. Tests have shown that reducing the .filling pressure will M e r increase the gamma-ray dose range in which the detectors can be operated.