skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Higher-Resolution Data-Dependent Selective External ION Accumulation for Capillary LC-FTICR

Journal Article · · International Journal of Mass Spectrometry, 218(3):265-279

Data-dependent selective external ion ejection with improved resolution is demonstrated with a 3.5 tesla FTICR instrument employing DREAMS (Dynamic Range Enhancement Applied to Mass Spectrometry) technology. To correct for the fringing rf-field aberrations each rod of the selection quadrupole has been segmented into three sections, so that ion excitation and ejection was performed by applying auxiliary rf-only waveforms in the region of the middle segments. Two different modes of external ion trapping and ejection were studied with the mixtures of model peptides and a tryptic digest of bovine serum albumin. A mass resolution of about 100 had been attained for rf-only dipolar ejection in a quadrupole operating at a Mathieu parameter q of ~0.45. LC-ESI-DREAMS-FTICR analysis of a 0.1 mg/mL solution of bovine serum albumin digest resulted in detection of 82 unique tryptic peptides with mass measurement errors lower than 5 ppm, providing 100 % sequence coverage of the protein.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
965637
Report Number(s):
PNNL-SA-38002; KP1501021; TRN: US200919%%656
Journal Information:
International Journal of Mass Spectrometry, 218(3):265-279, Vol. 218, Issue 3
Country of Publication:
United States
Language:
English