Accurate Characterization of the Shape of the HPGe Detector Peak Efficiency Curve for Application in PGNAA
Conference
·
OSTI ID:786213
Measurements in prompt gamma neutron activation analysis (PGNAA) typically have errors of the order of {+-}2 to 3%. A method is described that can yield a lower limit on the uncertainty in relative efficiencies applied to PGNAA of about {+-}0.1%.
- Research Organization:
- University of Cincinnati, Cincinnati, Ohio (US)
- Sponsoring Organization:
- None (US)
- OSTI ID:
- 786213
- Report Number(s):
- ISSN 0003-018X; CODEN TANSAO; ISSN 0003-018X; CODEN TANSAO; TRN: US0108838
- Resource Relation:
- Conference: 2001 Annual Meeting, Milwaukee, WI (US), 06/17/2001--06/21/2001; Other Information: Transactions of the American Nuclear Society, volume 84; PBD: 17 Jun 2001
- Country of Publication:
- United States
- Language:
- English
Similar Records
HPGe detector shielding optimization with MCNPX for the MEDINA PGNAA cell
Neutron Damage in Mechanically-Cooled High-Purity Germanium Detectors for Field-Portable Prompt Gamma Neutron Activation Analysis (PGNAA) Systems
Active interrogation of Special Nuclear Material containers using AmBe quasi-forward biased directional source and PGNAA
Conference
·
Wed Jul 01 00:00:00 EDT 2015
·
OSTI ID:786213
+2 more
Neutron Damage in Mechanically-Cooled High-Purity Germanium Detectors for Field-Portable Prompt Gamma Neutron Activation Analysis (PGNAA) Systems
Conference
·
Tue Oct 01 00:00:00 EDT 2013
·
OSTI ID:786213
+1 more
Active interrogation of Special Nuclear Material containers using AmBe quasi-forward biased directional source and PGNAA
Journal Article
·
Wed Nov 06 00:00:00 EST 2019
· Applied Radiation and Isotopes
·
OSTI ID:786213