skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Single-laboratory evaluation and modification of US EPA (Environmental Protection Agency) Methods 7470 and 7471 for the determination of mercury in aqueous and solid hazardous wastes (journal version)

Technical Report ·
OSTI ID:6368001

The current U.S. Environmental Protection Agency (U.S. EPA) protocols for mercury determinations in aqueous and solid-waste samples (SW-846 Methods 7470 and 7471) using recirculating cold-vapor atomic absorption spectrometry (CV-AAS) were evaluated. The U.S. EPA protocols have been modified in a single-laboratory study to facilitate additional QC measures, to enhance detectability for low-level mercury concentrations, and to eliminate nonspecific vapor-absorption interferences. Volumetric manipulations for additional QC measures, if required, are facilitated by performing the sample digestions in Erlenmeyer flasks rather than in the current Biochemical Oxygen Demand (BOD) reduction-aeration bottles. Instrument detectability is improved 10-fold by using a gas sparging bottle as a dedicated reduction-aeration vessel and a silver wool-amalgamation CV-AAS system operated in an open configuration. The on-line amalgamation thermal-desorption process of the modified CV-AAS system eliminates interfering water and organic matrix vapors prior to the mercury absorption measurement. Good accuracy and precision have been obtained with the amalgamation CV-AAS system for the analyses of four reference sediment materials.

Research Organization:
Battelle Columbus Div., OH (USA)
OSTI ID:
6368001
Report Number(s):
PB-89-130389/XAB
Resource Relation:
Other Information: Pub. in Environment International, Vol. 13, 475-489(1987)
Country of Publication:
United States
Language:
English