Title: 5-F catheter in cerebral angiography

Although the 5-F catheter is reputed to cause less vascular trauma than larger catheters, subintimal injections of contrast material have occurred following intimal damage by the catheter tip. Microscopic studies of the tips of two widely used 5-F polyethylene catheters have revealed a difference in configuration resulting in one of the catheters becoming markedly damaged during angiography. The authors make recommendations for finishing and protecting the catheter tip.
Authors:
; ; ;
Publication Date:
OSTI Identifier:
5799471
Resource Type:
Journal Article
Resource Relation:
Journal Name: Radiology; (United States); Journal Volume: 141:2
Research Org:
Harvard Medical School, Boston, MA
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; BLOOD VESSELS; BIOMEDICAL RADIOGRAPHY; CEREBELLUM; DIAGNOSIS; SURGERY; BODY; BRAIN; CARDIOVASCULAR SYSTEM; CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; MEDICINE; NERVOUS SYSTEM; NUCLEAR MEDICINE; ORGANS; RADIOLOGY 550602* -- Medicine-- External Radiation in Diagnostics-- (1980-)