Troubleshooting the General Electric Company Neutron Devices Department's deionized water system
In May 1982, the General Electric Neutron Devices Department's deionized (DI) water system was found to be highly contaminated with fresh water roundworms which caused severe filter plugging problems. Subsequently, a DI water improvement program was developed by GEND's Process Technology, Contamination Control, and Facilities Engineering groups. This program included removal of dead ends in the distribution system, a consultant's review, significant modification of the DI water system, modification of system operation, design of a quality plan, and changes in the types of water analysis. During implementation of the improvement program, a severe bacteria contamination problem occurred due to a contractor accident. Correction of this problem required sterilization of portions of the DI water system and significant modification of the water filtration system. The system modifications and sterilization procedures have significantly reduced bacteria and total solids contamination while resistivity is generally increasing.
- Research Organization:
- General Electric Co., St. Petersburg, FL (USA). Neutron Devices Dept.
- DOE Contract Number:
- AC04-76DP00656
- OSTI ID:
- 5696326
- Report Number(s):
- GEPP-TIS-761; ON: DE84002960; TRN: 83-027466
- Country of Publication:
- United States
- Language:
- English
Similar Records
Producing Light Oil from a Frozen Reservoir: Reservoir and Fluid Characterization of Umiat Field, National Petroleum Reserve, Alaska
High Ion-Accessible Surface Area CNT-Ultracapacitors for Groundwater Desalination
Related Subjects
37 INORGANIC
ORGANIC
PHYSICAL AND ANALYTICAL CHEMISTRY
WATER
PURIFICATION
BACTERIA
FILTERS
LABORATORY EQUIPMENT
NEMATODES
PLANNING
QUALITY ASSURANCE
STERILIZATION
ASCHELMINTHES
EQUIPMENT
HELMINTHS
HYDROGEN COMPOUNDS
MICROORGANISMS
OXYGEN COMPOUNDS
070200* - Radiation Sources
400105 - Separation Procedures