Title: Laser videofluorometer system for real-time characterization of high-performance liquid chromatographic eluate. [3-hydroxy-benzo(a)pyrene]

Journal Article · · Anal. Chem.; (United States)
DOI:https://doi.org/10.1021/ac00126a054· OSTI ID:5109501

A second generation videofluorometer has been developed for real-time characterization of high-performance liquid chromatographic eluate. The instrument features a nitrogen-laser-pumped dye laser as excitation source and quarter meter polychromator/microchannel plate-intensified diode array as fluorescence detector. The dye laser cavity is tuned with a moving-iron galvanometer scanner grating drive, permitting the laser output to be changed to any wavelength in its range in less than 40 ms. Thus, the optimum excitation wavelength can be chosen for each chromatographic region. A minimum detection limit of 13 pptr has been obtained for 3-hydroxy-benzo(a)pyrene in a conventional fluorescence cuvette with a 30-s data acquisition. For the same substance eluted chromatographically, a minimum detection limit of 50 pg has been obtained, and a linear dynamic range of greater than 3 orders of magnitude observed. An extract of soil that had been contaminated with polyaromatic hydrocarbons was analyzed as a practical test of the system, permitting the quantitation of three known species, and the identification and quantitation of a previously unknown fourth compound.

Research Organization:
Univ. of Washington, Seattle
OSTI ID:
5109501
Journal Information:
Anal. Chem.; (United States), Vol. 58:13
Country of Publication:
United States
Language:
English