Title: Radioimmunoassay of ''free thyroxin'' in dried blood spots on filter paper - preliminary observations on the effective differentiation of subjects with congenital hypothyroidism from those with subnormal thyroxin-binding globulin and normal subjects

Journal Article · · Clin. Chem. (Winston-Salem, N.C.); (United States)
OSTI ID:5031520

In this sensitive, simple method for measuring ''free thyroxin'' (FT/sub 4/) in eluates of dried blood spots on filter paper by use of a radioimmunoassay kit (Amerlex Free T/sub 4/ RIA), the measurable range of FT/sub 4/ is 1.8 to 57 ng/L (equivalent to the concentration in serum), or 7 to 237 fg/tube. The mean coefficients of variation for within assay-within spots, within assay-between spots, and between assays were 5.3%, 5.0%, and 6.2%, respectively. FT/sub 4/ in blood spotted on filter paper is stable for at least a month when dried and kept at either -20/sup 0/C, 4/sup 0/C, room temperature (about 25/sup 0/C), or 37/sup 0/C. The results for FT/sub 4/ in dried blood spots correlated closely with the free-T/sub 4/ concentration in serum (r = 0.99). The method can be used to differentiate cases of primary and secondary hypothyroidism from normal subjects and those with subnormal thyroxin-binding globulin. This method may be useful in screening for congenital hypothyroidism, because sample-retesting is not necessary.

Research Organization:
Osaka Univ., Japan
OSTI ID:
5031520
Journal Information:
Clin. Chem. (Winston-Salem, N.C.); (United States), Vol. 28:3
Country of Publication:
United States
Language:
English