Title: Nitrogen dioxide and kerosene-flame soot calibration of photoacoustic instruments for measurement of light absorption by aerosols

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1322585· OSTI ID:40204900

A nitrogen dioxide calibration method is developed to evaluate the theoretical calibration for a photoacoustic instrument used to measure light absorption by atmospheric aerosols at a laser wavelength of 532.0 nm. This method uses high concentrations of nitrogen dioxide so that both a simple extinction and the photoacoustically obtained absorption measurement may be performed simultaneously. Since Rayleigh scattering is much less than absorption for the gas, the agreement between the extinction and absorption coefficients can be used to evaluate the theoretical calibration, so that the laser gas spectra are not needed. Photoacoustic theory is developed to account for strong absorption of the laser beam power in passage through the resonator. Findings are that the photoacoustic absorption based on heat-balance theory for the instrument compares well with absorption inferred from the extinction measurement, and that both are well within values represented by published spectra of nitrogen dioxide. Photodissociation of nitrogen dioxide limits the calibration method to wavelengths longer than 398 nm. Extinction and absorption at 532 and 1047 nm were measured for kerosene-flame soot to evaluate the calibration method, and the single scattering albedo was found to be 0.31 and 0.20 at these wavelengths, respectively.

Sponsoring Organization:
(US)
OSTI ID:
40204900
Journal Information:
Review of Scientific Instruments, Vol. 71, Issue 12; Other Information: DOI: 10.1063/1.1322585; Othernumber: RSINAK000071000012004545000001; 035012RSI; PBD: Dec 2000; ISSN 0034-6748
Publisher:
The American Physical Society
Country of Publication:
United States
Language:
English