Determination of BTX in air by hand-held gas chromatography
Conference
·
OSTI ID:230966
- Amoco Corp., Naperville, IL (United States)
A hand-held gas chromatograph (Photovac Snapshot GC) has been validated for field determination of low to sub-ppmv concentrations of BTX in air. This instrument carries a Class 1 Division 1, Groups A,B,C,D safety rating and performs a benzene-only analysis in 4 minutes and a benzene, toluene and C8 aromatics (i.e. ethylbenzene, o,m,p-xylenes) analysis in 10 minutes. The detection limit for benzene is 0.1 ppmv and 1.0 ppmv for toluene and C8 aromatics. The Snapshot GC has been successfully utilized by field personnel for several field applications, such as Industrial Hygiene area monitoring, vapor recovery unit carbon bed outlet measurements and remediation system vent measurements.
- OSTI ID:
- 230966
- Report Number(s):
- CONF-950209-; TRN: IM9623%%276
- Resource Relation:
- Conference: 4. international symposium on field screening methods for hazardous wastes and toxic chemicals, Las Vegas, NV (United States), 22-24 Feb 1995; Other Information: PBD: 1995; Related Information: Is Part Of Field screening methods for hazardous wastes and toxic chemicals. VIP-47, Volume 1; PB: 733 p.
- Country of Publication:
- United States
- Language:
- English
Similar Records
Fast gas chromatography for industrial hygiene analysis and monitoring
Superfund innovative technology evaluation (site) program evaluation report for antox BTX water screen (BTX immunoassay)
Field gas chromatography analysis as input for vadose zone modeling: A case study
Miscellaneous
·
Sun Jan 01 00:00:00 EST 1989
·
OSTI ID:230966
Superfund innovative technology evaluation (site) program evaluation report for antox BTX water screen (BTX immunoassay)
Technical Report
·
Tue Jun 01 00:00:00 EDT 1993
·
OSTI ID:230966
+1 more
Field gas chromatography analysis as input for vadose zone modeling: A case study
Conference
·
Sun Dec 31 00:00:00 EST 1995
·
OSTI ID:230966