Title: SU-D-209-02: Percent Depth Dose Curves for Fluoroscopic X-Ray Beam Qualities Incorporating Copper Filtration

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4955663· OSTI ID:22624407
 [1]; ; ;  [1];  [2]
  1. Cleveland Clinic Foundation, Cleveland, OH (United States)
  2. Wayne State University School of Medicine, Detroit, MI (United States)

Purpose: The purpose of this investigation was to quantify percent depth dose (PDD) curves for fluoroscopic x-ray beam qualities incorporating added copper filtration. Methods: A PTW (Freiburg, Germany) MP3 water tank was used with a Standard Imaging (Middleton, WI) Exradin Model 11 Spokas Chamber to measure PDD curves for 60, 80, 100 and 120 kVp x-ray beams with copper filtration ranging from 0.0–0.9 mm at 22cm and 42cm fields of view from 0 to 150 mm of water. A free-in-air monitor chamber was used to normalize the water tank data to fluctuations in output from the fluoroscope. The measurements were acquired on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope. The fluoroscope was inverted from the typical orientation providing an x-ray beam originating from above the water tank. The water tank was positioned so that the water level was located at 60cm from the focal spot; which also represents the focal spot to interventional reference plane distance for that fluoroscope. Results: PDDs for 60, 80, 100, and 120 kVp with 0 mm of copper filtration compared well to previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)] for those beam qualities given differences in fluoroscopes, geometric orientation, type of ionization chamber, and the water tank used for data collection. PDDs for 60, 80, 100, and 120 kVp with copper filtration were obtained and are presented, which have not been previously investigated and published. Conclusion: The equipment and processes used to acquire the reported data were sound and compared well with previously published data for PDDs without copper filtration. PDD data for the fluoroscopic x-ray beams incorporating copper filtration can be used as reference data for estimating organ or soft tissue dose at depth involving similar beam qualities or for comparison with mathematical models.

OSTI ID:
22624407
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English