Title: SU-C-BRB-06: Dosimetric Impact of Breast Contour Reconstruction Errors in GammaPod Stereotactic Radiotherapy

Abstract

Purpose: The first GammaPod™ unit, a dedicated prone stereotactic treatment device for early stage breast cancer, has been installed and commissioned at University of Maryland School of Medicine. The objective of this study was to investigate potential dosimetric impact of inaccurate breast contour. Methods: In GammaPod treatments, patient’s beast is immobilized by a breast cup device (BCID) throughout the entire same-day imaging and treatment procedure. 28 different BICD sizes are available to accommodate patients with varying breast sizes. A mild suction helps breast tissue to conform to the shape of the cup with selected size. In treatment planning, dose calculation utilizes previously calculated dose distributions for available cup geometry rather than the breast shape from CT image. Patient CT images with breast cups indicate minor geometric discrepancy between the matched shape of the cup and the breast contour, i.e., the contour size is larger or smaller. In order to investigate the dosimetric impact of these discrepancies, we simulated such discrepancies and reassessed the dose to target as well as skin. Results: In vicinity of skin, hot/cold spots were found when matched cup size was smaller/larger than patient’s breast after comparing the corrected dose profiles from Monte Carlo simulation with themore » planned dose from TPS. The overdosing/underdosing of target could yield point dose differences as large as 5% due to these setup errors (D95 changes within 2.5%). Maximal skin dose was overestimated/underestimated up to 25%/45% when matched cup size was larger/smaller than real breast contour. Conclusion: The dosimetric evaluation suggests substantial underdosing/overdosing with inaccurate cup geometry during planning, which is acceptable for current clinical trial. Further studies are needed to evaluate such impact to treating small volume close to skin.« less

Authors:
 [1]; ;  [2];  [1]
  1. Xcision Medical Systems LLC, Columbia, MD (United States)
  2. University Maryland School of Medicine, Baltimore, MD (United States)
Publication Date:
OSTI Identifier:
22624323
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; CLINICAL TRIALS; COMPUTERIZED SIMULATION; COMPUTERIZED TOMOGRAPHY; ERRORS; MAMMARY GLANDS; MONTE CARLO METHOD; NEOPLASMS; PATIENTS; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY

Citation Formats

Niu, Y, Becker, S, Mutaf, Y, Yu, C, and University Maryland School of Medicine, Baltimore, MD. SU-C-BRB-06: Dosimetric Impact of Breast Contour Reconstruction Errors in GammaPod Stereotactic Radiotherapy. United States: N. p., 2016. Web. doi:10.1118/1.4955560.
Niu, Y, Becker, S, Mutaf, Y, Yu, C, & University Maryland School of Medicine, Baltimore, MD. SU-C-BRB-06: Dosimetric Impact of Breast Contour Reconstruction Errors in GammaPod Stereotactic Radiotherapy. United States. https://doi.org/10.1118/1.4955560
Niu, Y, Becker, S, Mutaf, Y, Yu, C, and University Maryland School of Medicine, Baltimore, MD. 2016. "SU-C-BRB-06: Dosimetric Impact of Breast Contour Reconstruction Errors in GammaPod Stereotactic Radiotherapy". United States. https://doi.org/10.1118/1.4955560.
@article{osti_22624323,
title = {SU-C-BRB-06: Dosimetric Impact of Breast Contour Reconstruction Errors in GammaPod Stereotactic Radiotherapy},
author = {Niu, Y and Becker, S and Mutaf, Y and Yu, C and University Maryland School of Medicine, Baltimore, MD},
abstractNote = {Purpose: The first GammaPod™ unit, a dedicated prone stereotactic treatment device for early stage breast cancer, has been installed and commissioned at University of Maryland School of Medicine. The objective of this study was to investigate potential dosimetric impact of inaccurate breast contour. Methods: In GammaPod treatments, patient’s beast is immobilized by a breast cup device (BCID) throughout the entire same-day imaging and treatment procedure. 28 different BICD sizes are available to accommodate patients with varying breast sizes. A mild suction helps breast tissue to conform to the shape of the cup with selected size. In treatment planning, dose calculation utilizes previously calculated dose distributions for available cup geometry rather than the breast shape from CT image. Patient CT images with breast cups indicate minor geometric discrepancy between the matched shape of the cup and the breast contour, i.e., the contour size is larger or smaller. In order to investigate the dosimetric impact of these discrepancies, we simulated such discrepancies and reassessed the dose to target as well as skin. Results: In vicinity of skin, hot/cold spots were found when matched cup size was smaller/larger than patient’s breast after comparing the corrected dose profiles from Monte Carlo simulation with the planned dose from TPS. The overdosing/underdosing of target could yield point dose differences as large as 5% due to these setup errors (D95 changes within 2.5%). Maximal skin dose was overestimated/underestimated up to 25%/45% when matched cup size was larger/smaller than real breast contour. Conclusion: The dosimetric evaluation suggests substantial underdosing/overdosing with inaccurate cup geometry during planning, which is acceptable for current clinical trial. Further studies are needed to evaluate such impact to treating small volume close to skin.},
doi = {10.1118/1.4955560},
url = {https://www.osti.gov/biblio/22624323}, journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}