Title: A novel calibration method for non-orthogonal shaft laser theodolite measurement system

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4943017· OSTI ID:22597099
; ;  [1];  [2]
  1. State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072 (China)
  2. College of Electrical Engineering and Automation, Tianjin Key Laboratory of Process Measurement and Control, Tianjin University, Tianjin 300072 (China)

Non-orthogonal shaft laser theodolite (N-theodolite) is a new kind of large-scale metrological instrument made up by two rotary tables and one collimated laser. There are three axes for an N-theodolite. According to naming conventions in traditional theodolite, rotary axes of two rotary tables are called as horizontal axis and vertical axis, respectively, and the collimated laser beam is named as sight axis. And the difference between N-theodolite and traditional theodolite is obvious, since the former one with no orthogonal and intersecting accuracy requirements. So the calibration method for traditional theodolite is no longer suitable for N-theodolite, while the calibration method applied currently is really complicated. Thus this paper introduces a novel calibration method for non-orthogonal shaft laser theodolite measurement system to simplify the procedure and to improve the calibration accuracy. A simple two-step process, calibration for intrinsic parameters and for extrinsic parameters, is proposed by the novel method. And experiments have shown its efficiency and accuracy.

OSTI ID:
22597099
Journal Information:
Review of Scientific Instruments, Vol. 87, Issue 3; Other Information: (c) 2016 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English