End loss analyzer system for measurements of plasma flux at the C-2U divertor electrode
- Tri Alpha Energy, P.O. Box 7010, Rancho Santa Margarita, California 92688 (United States)
An end loss analyzer system consisting of electrostatic, gridded retarding-potential analyzers and pyroelectric crystal bolometers was developed to characterize the plasma loss along open field lines to the divertors of C-2U. The system measures the current and energy distribution of escaping ions as well as the total power flux to enable calculation of the energy lost per escaping electron/ion pair. Special care was taken in the construction of the analyzer elements so that they can be directly mounted to the divertor electrode. An attenuation plate at the entrance to the gridded retarding-potential analyzer reduces plasma density by a factor of 60 to prevent space charge limitations inside the device, without sacrificing its angular acceptance of ions. In addition, all of the electronics for the measurement are isolated from ground so that they can float to the bias potential of the electrode, 2 kV below ground.
- OSTI ID:
- 22596487
- Journal Information:
- Review of Scientific Instruments, Vol. 87, Issue 11; Other Information: (c) 2016 Author(s); Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Double parallel-plate electrostatic analyzer for Tara end-loss studies
Large-acceptance-angle gridded analyzers in an axial magnetic field