Title: MO-DE-210-05: Improved Accuracy of Liver Feature Motion Estimation in B-Mode Ultrasound for Image-Guided Radiation Therapy

Abstract

Purpose: In similarity-measure based motion estimation incremental tracking (or template update) is challenging due to quantization, bias and accumulation of tracking errors. A method is presented which aims to improve the accuracy of incrementally tracked liver feature motion in long ultrasound sequences. Methods: Liver ultrasound data from five healthy volunteers under free breathing were used (15 to 17 Hz imaging rate, 2.9 to 5.5 minutes in length). A normalised cross-correlation template matching algorithm was implemented to estimate tissue motion. Blood vessel motion was manually annotated for comparison with three tracking code implementations: (i) naive incremental tracking (IT), (ii) IT plus a similarity threshold (ST) template-update method and (iii) ST coupled with a prediction-based state observer, known as the alpha-beta filter (ABST). Results: The ABST method produced substantial improvements in vessel tracking accuracy for two-dimensional vessel motion ranging from 7.9 mm to 40.4 mm (with mean respiratory period: 4.0 ± 1.1 s). The mean and 95% tracking errors were 1.6 mm and 1.4 mm, respectively (compared to 6.2 mm and 9.1 mm, respectively for naive incremental tracking). Conclusions: High confidence in the output motion estimation data is required for ultrasound-based motion estimation for radiation therapy beam tracking and gating. The methodmore » presented has potential for monitoring liver vessel translational motion in high frame rate B-mode data with the required accuracy. This work is support by Cancer Research UK Programme Grant C33589/A19727.« less

Authors:
; ;  [1]
  1. The Institute of Cancer Research & Royal Marsden, Sutton and London (United Kingdom)
Publication Date:
OSTI Identifier:
22562874
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ACCURACY; ALGORITHMS; ANIMAL TISSUES; BIOMEDICAL RADIOGRAPHY; BLOOD; BLOOD VESSELS; COMPARATIVE EVALUATIONS; ERRORS; IMAGES; LIVER; NEOPLASMS; QUANTIZATION; RADIOTHERAPY; RESPIRATION

Citation Formats

O’Shea, T, Bamber, J, and Harris, E. MO-DE-210-05: Improved Accuracy of Liver Feature Motion Estimation in B-Mode Ultrasound for Image-Guided Radiation Therapy. United States: N. p., 2015. Web. doi:10.1118/1.4925367.
O’Shea, T, Bamber, J, & Harris, E. MO-DE-210-05: Improved Accuracy of Liver Feature Motion Estimation in B-Mode Ultrasound for Image-Guided Radiation Therapy. United States. https://doi.org/10.1118/1.4925367
O’Shea, T, Bamber, J, and Harris, E. 2015. "MO-DE-210-05: Improved Accuracy of Liver Feature Motion Estimation in B-Mode Ultrasound for Image-Guided Radiation Therapy". United States. https://doi.org/10.1118/1.4925367.
@article{osti_22562874,
title = {MO-DE-210-05: Improved Accuracy of Liver Feature Motion Estimation in B-Mode Ultrasound for Image-Guided Radiation Therapy},
author = {O’Shea, T and Bamber, J and Harris, E},
abstractNote = {Purpose: In similarity-measure based motion estimation incremental tracking (or template update) is challenging due to quantization, bias and accumulation of tracking errors. A method is presented which aims to improve the accuracy of incrementally tracked liver feature motion in long ultrasound sequences. Methods: Liver ultrasound data from five healthy volunteers under free breathing were used (15 to 17 Hz imaging rate, 2.9 to 5.5 minutes in length). A normalised cross-correlation template matching algorithm was implemented to estimate tissue motion. Blood vessel motion was manually annotated for comparison with three tracking code implementations: (i) naive incremental tracking (IT), (ii) IT plus a similarity threshold (ST) template-update method and (iii) ST coupled with a prediction-based state observer, known as the alpha-beta filter (ABST). Results: The ABST method produced substantial improvements in vessel tracking accuracy for two-dimensional vessel motion ranging from 7.9 mm to 40.4 mm (with mean respiratory period: 4.0 ± 1.1 s). The mean and 95% tracking errors were 1.6 mm and 1.4 mm, respectively (compared to 6.2 mm and 9.1 mm, respectively for naive incremental tracking). Conclusions: High confidence in the output motion estimation data is required for ultrasound-based motion estimation for radiation therapy beam tracking and gating. The method presented has potential for monitoring liver vessel translational motion in high frame rate B-mode data with the required accuracy. This work is support by Cancer Research UK Programme Grant C33589/A19727.},
doi = {10.1118/1.4925367},
url = {https://www.osti.gov/biblio/22562874}, journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 42,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2015},
month = {Mon Jun 15 00:00:00 EDT 2015}
}