Title: SU-E-T-365: Estimation of Neutron Ambient Dose Equivalents for Radioprotection Exposed Workers in Radiotherapy Facilities Based On Characterization Patient Risk Estimation

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924726· OSTI ID:22548415
; ;  [1]; ;  [2];  [3];  [4];  [5]
  1. Departamento de Fisiologia Medica y Biofisica, Universidad de Sevilla (Spain)
  2. Departament de Fisica, Universitat Autonoma de Barcelona, Bellaterra (Spain)
  3. Health and Safety Department, ALBA Synchrotron Light Source, Cerdanyola del Valles (Spain)
  4. Instituto de Fisica, Pontificia Universidad Catolica de Chile, Santiago (Chile)
  5. Laboratori Nazionali di Frascati, Istituto Nazionale di Fisica Nucleare (INFN) (Italy)

Purpose: Previous measurements with Bonner spheres{sup 1} showed that normalized neutron spectra are equal for the majority of the existing linacs{sup 2}. This information, in addition to thermal neutron fluences obtained in the characterization procedure{sup 3}3, would allow to estimate neutron doses accidentally received by exposed workers, without the need of an extra experimental measurement. Methods: Monte Carlo (MC) simulations demonstrated that the thermal neutron fluence distribution inside the bunker is quite uniform, as a consequence of multiple scatter in the walls{sup 4}. Although inverse square law is approximately valid for the fast component, a more precise calculation could be obtained with a generic fast fluence distribution map around the linac, from MC simulations{sup 4}. Thus, measurements of thermal neutron fluences performed during the characterization procedure{sup 3}, together with a generic unitary spectra{sup 2}, would allow to estimate the total neutron fluences and H*(10) at any point{sup 5}. As an example, we compared estimations with Bonner sphere measurements{sup 1}, for two points in five facilities: 3 Siemens (15–23 MV), Elekta (15 MV) and Varian (15 MV). Results: Thermal neutron fluences obtained from characterization, are within (0.2–1.6×10{sup 6}) cm−{sup 2}•Gy{sup −1} for the five studied facilities. This implies ambient equivalent doses ranging from (0.27–2.01) mSv/Gy 50 cm far from the isocenter and (0.03–0.26) mSv/Gy at detector location with an average deviation of ±12.1% respect to Bonner measurements. Conclusion: The good results obtained demonstrate that neutron fluence and H*(10) can be estimated based on: (a) characterization procedure established for patient risk estimation in each facility, (b) generic unitary neutron spectrum and (c) generic MC map distribution of the fast component. [1] Radiat. Meas (2010) 45: 1391 – 1397; [2] Phys. Med. Biol (2012) 5 7:6167–6191; [3] Med. Phys (2015) 42:276 - 281. [4] IFMBE (2012) 39: 1245–1248. [5] ICRU Report 57 (1998)

OSTI ID:
22548415
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English