SU-E-T-234: Daily Quality Assurance for a Six Degrees of Freedom Couch Using a Novel Phantom
- Ohio State Univ, Columbus, OH (United States)
- Memorial Medical Center, Modesto, CA (United States)
Purpose: To test the accuracy and reproducibility of both translational and rotational movements for a couch with six degrees of freedom (6DoF) using a novel phantom design Methods: An end-to-end test was carried out using two different phantoms. A 6 cm3 cube with a central fiducial BB (WL-QA Sun Nuclear) and a custom fabricated rectangular prism (31 cm x 8 cm x 8 cm), placed on a baseplate with known angular offsets for pitch, roll and yaw with a central fiducial BB and unique surface structures for registration purposes, were used. The end-to-end test included an initial CT simulation for a reference study, setup to an offset mark on each phantom, registration of the reference CT to the acquired cone-beam CT, and final Winston-Lutz delivery at four cardinal gantry angles. Results for both translational and rotational movements were recorded and compared for both phantoms. Results: Translational and rotational measurements were performed with a PerfectPitch (Varian) couch for 10 trials for both phantoms. Distinct translational shifts were [−5.372±0.384mm, −10.183±0.137mm, 14.028±0.155mm] for the cube and [7.520±0.159mm, −9.117±0.101mm, 16.273±0.115mm] for the prototype phantom for lateral, longitudinal, and vertical shifts, respectively. Distinct rotational adjustments were [1.121±0.102o, −1.067±0.235o, −2.662±0.380o] for the cube and [2.534±0.059o, 1.994±0.025o, 2.094±0.076o] for the prototype for pitch, roll, and yaw, respectively. Winston-Lutz test results performed after 6DoF couch correction from each cardinal gantry angle ranged from 0.26–0.72mm for the cube and 0.55–0.86mm for the prototype. Conclusion: The prototype phantom is more precise for both translational and rotational adjustments compared to a commercial phantom. The design of the prototype phantom allows for a more discernible visual confirmation of correct translational and rotational adjustments with the prototype phantom. Winston-Lutz results are more accurate for the commercial phantom but are still within tolerance for the prototype phantom.
- OSTI ID:
- 22548292
- Journal Information:
- Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
MO-F-CAMPUS-J-01: A Novel and Efficient Daily QA Program for Both Modern Linear Accelerator and Optical Surface Motion Systems
SU-F-J-177: A Novel Image Analysis Technique (center Pixel Method) to Quantify End-To-End Tests