SU-E-T-195: Commissioning the Neutron Production of a Varian TrueBeam Linac
Abstract
Purpose: The purpose of this work is the characterization of a new Varian TrueBeam™ facility in terms of neutron production, in order to estimate neutron equivalent dose in organs during radiotherapy treatments. Methods: The existing methodology [1] was used with the reference SRAMnd detector, calibrated in terms of thermal neutron fluence at the reference field operated by PTB (Physikalisch-Technische-Bundesanstalt) at the GeNF (Geesthacht-Neutron-Facility) with the GKSS reactor FRG-1 [2]. Thermal neutron fluence for the 5 available possibilities was evaluated: 15 MV and 10&6 MV with and without Flattening Filter (FF and FFF, respectively). Irradiation conditions are as described in [3]. In addition, three different collimator-MLC configurations were studied for 15 MV: (a) collimator of 10×10 cm{sup 2} and MLC fully retracted (reference), (b) field sizes of 20×20 cm{sup 2} and 10×10 cm{sup 2} for collimator and MLC respectively, and (c) collimator and MLC aperture of 10×10 cm{sup 2}. Results: Thermal fluence rate at the “reference point” [3], as a consequence of the neutron production, obtained for (a) conformation in 15 MV is (1.45±0.11) x10{sup 4} n•cm{sup 2}/MU. Configurations (b) and (c) gave fluences of 96.6% and 97.8% of the reference (a). Neutron production decreases up to 8.6% and 5.7% formore »
- Authors:
- Publication Date:
- OSTI Identifier:
- 22545318
- Resource Type:
- Journal Article
- Journal Name:
- Medical Physics
- Additional Journal Information:
- Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
- Subject:
- 07 ISOTOPES AND RADIATION SOURCES; 60 APPLIED LIFE SCIENCES; APERTURES; BEAMS; COLLIMATORS; COMMISSIONING; CONFIGURATION; DOSE EQUIVALENTS; FRG-1 REACTOR; IRRADIATION; LINEAR ACCELERATORS; NEUTRON FLUENCE; ORGANS; RADIOTHERAPY; THERMAL NEUTRONS
Citation Formats
Irazola, L, Brualla, L, Rosello, J, Terron, JA, Sanchez-Nieto, B, Bedogni, R, and Sanchez-Doblado, F. SU-E-T-195: Commissioning the Neutron Production of a Varian TrueBeam Linac. United States: N. p., 2015.
Web. doi:10.1118/1.4924556.
Irazola, L, Brualla, L, Rosello, J, Terron, JA, Sanchez-Nieto, B, Bedogni, R, & Sanchez-Doblado, F. SU-E-T-195: Commissioning the Neutron Production of a Varian TrueBeam Linac. United States. https://doi.org/10.1118/1.4924556
Irazola, L, Brualla, L, Rosello, J, Terron, JA, Sanchez-Nieto, B, Bedogni, R, and Sanchez-Doblado, F. 2015.
"SU-E-T-195: Commissioning the Neutron Production of a Varian TrueBeam Linac". United States. https://doi.org/10.1118/1.4924556.
@article{osti_22545318,
title = {SU-E-T-195: Commissioning the Neutron Production of a Varian TrueBeam Linac},
author = {Irazola, L and Brualla, L and Rosello, J and Terron, JA and Sanchez-Nieto, B and Bedogni, R and Sanchez-Doblado, F},
abstractNote = {Purpose: The purpose of this work is the characterization of a new Varian TrueBeam™ facility in terms of neutron production, in order to estimate neutron equivalent dose in organs during radiotherapy treatments. Methods: The existing methodology [1] was used with the reference SRAMnd detector, calibrated in terms of thermal neutron fluence at the reference field operated by PTB (Physikalisch-Technische-Bundesanstalt) at the GeNF (Geesthacht-Neutron-Facility) with the GKSS reactor FRG-1 [2]. Thermal neutron fluence for the 5 available possibilities was evaluated: 15 MV and 10&6 MV with and without Flattening Filter (FF and FFF, respectively). Irradiation conditions are as described in [3]. In addition, three different collimator-MLC configurations were studied for 15 MV: (a) collimator of 10×10 cm{sup 2} and MLC fully retracted (reference), (b) field sizes of 20×20 cm{sup 2} and 10×10 cm{sup 2} for collimator and MLC respectively, and (c) collimator and MLC aperture of 10×10 cm{sup 2}. Results: Thermal fluence rate at the “reference point” [3], as a consequence of the neutron production, obtained for (a) conformation in 15 MV is (1.45±0.11) x10{sup 4} n•cm{sup 2}/MU. Configurations (b) and (c) gave fluences of 96.6% and 97.8% of the reference (a). Neutron production decreases up to 8.6% and 5.7% for the 10 MV FF and FFF beams, respectively. Finally, it decreases up to 2.8% and 0.1% for the 6 MV FF and FFF modes, respectively. Conclusion: This work evaluates thermal neutron production of Varian TrueBeam™ system for organ equivalent dose estimation. The small difference in collimator-MLC configuration shows the universality of the methodology [3]. A decrease in this production is shown when decreasing energy from 15 to 10 MV and an almost negligible production was found for 6 MV. Moreover, a lower neutron contribution is observed for the FFF modes.[1]Phys Med Biol,2012;57:6167–6191.[2]Radiat Meas,2010;45:1513–1517.[3]Med Phys,2015;42:276–281.},
doi = {10.1118/1.4924556},
url = {https://www.osti.gov/biblio/22545318},
journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 42,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2015},
month = {Mon Jun 15 00:00:00 EDT 2015}
}