skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-139: Automated Daily EPID Exit Dose Analysis Uncovers Treatment Variations

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924501· OSTI ID:22545265
 [1]
  1. University of Southern California, Los Angeles, CA (United States)

Purpose: To evaluate a fully automated EPID exit dose system for its ability to detect daily treatment deviations including patient setup, delivery, and anatomy changes. Methods: PerFRACTION (Sun Nuclear Corporation) software is a system that uses integrated EPID images taken during patient treatment and automatically pulled from the Aria database and analyzed based on user-defined comparisons. This was used to monitor 20 plans consisting of a total of 859 fields for 18 patients, for a total of 251 fractions. Nine VMAT, 5 IMRT, and 6 3D plans were monitored. The Gamma analysis was performed for each field within a plan, comparing the first fraction against each of the other fractions in each treatment course. A 2% dose difference, 1 mm distance-to-agreement, and 10% dose threshold was used. These tight tolerances were chosen to achieve a high sensitivity to treatment variations. The field passed if 93% of the pixels had a Gamma of 1 or less. Results: Twenty-nine percent of the fields failed. The average plan passing rate was 92.5%.The average 3D plan passing rate was less than for VMAT or IMRT, 84%, vs. an average of 96.2%. When fields failed, an investigation revealed changes in patient anatomy or setup variations, often also leading to variations of transmission through immobilization devices. Conclusion: PerFRACTION is a fully automated system for determining daily changes in dose transmission through the patient that requires no effort other than for the imager panel to be deployed during treatment. A surprising number of fields failed the analysis and can be attributed to important treatment variations that would otherwise not be appreciated. Further study of inter-fraction treatment variations is possible and warranted. Sun Nuclear Corporation provided a license to the software described.

OSTI ID:
22545265
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English