Title: SU-E-J-176: Characterization of Inter-Fraction Breast Variability and the Implications On Delivered Dose

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924261· OSTI ID:22499283
; ; ; ;  [1]
  1. University of Cincinnati, Cincinnati, OH (United States)

Purpose: To systematically characterize inter-fraction breast variability and determine implications on delivered dose. Methods: Weekly port films were used to characterize breast setup variability. Five evenly spaced representative positions across the contour of each breast were chosen on the electronic port film in reference to graticule, and window and level was set such that the skin surface of the breast was visible. Measurements from the skin surface to treatment field edge were taken on each port film at each position and compared to the planning DRR, quantifying the variability. The systematic measurement technique was repeated for all port films for 20 recently treated breast cancer patients. Measured setup variability for each patient was modeled as a normal distribution. The distribution was randomly sampled from the model and applied as isocentric shifts in the treatment planning computer, representing setup variability for each fraction. Dose was calculated for each shifted fraction and summed to obtain DVHs and BEDs that modeled the dose with daily setup variability. Patients were categorized in to relevant groupings that were chosen to investigate the rigorousness of immobilization types, treatment techniques, and inherent anatomical difficulties. Mean position differences and dosimetric differences were evaluated between planned and delivered doses. Results: The setup variability was found to follow a normal distribution with mean position differences between the DRR and port film between − 8.6–3.5 mm with sigma range of 5.3–9.8 mm. Setup position was not found to be significantly different than zero. The mean seroma or whole breast PTV dosimetric difference, calculated as BED, ranged from a −0.23 to +1.13Gy. Conclusion: A systematic technique to quantify and model setup variability was used to calculate the dose in 20 breast cancer patients including variable setup. No statistically significant PTV or OAR BED differences were found between immobilization, treatment or anatomical differences investigated.

OSTI ID:
22499283
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English