skip to main content

Title: SU-E-T-592: OSL Response of Al2O3:C Detectors Exposed to Therapeutic Proton Beams

Purpose: To characterize the response of Al{sub 2}O{sub 3}:C optically stimulated luminescence (OSL) detectors (OSLDs) exposed to therapeutic proton beams of differing beam quality. Methods: We prepared Al{sub 2}O{sub 3}:C OSLDs from the same material as commercially available nanoDot dosimeters (Landauer, Inc). We irradiated the OSLDs in modulated proton beams of varying quality, as defined by the residual range. An absorbed dose to water of 0.2 Gy was delivered to all OSLDs with the residual range values varying from 0.5 to 23.5 cm (average LET in water from ∼0.5 to 2.5 keV/µm). To investigate the beam quality dependence of different emission bands within the OSL spectrum, we performed OSLD readouts using both continuous-wave stimulation (CW-OSL) and pulsed stimulation (P-OSL) with two sets of optical filters (Hoya U-340 and Kopp 5113). For all readout modes, the relative absorbed dose sensitivity (S{sub rel}) for each beam quality was calculated using OSLDs irradiated in a 6 MV photon beam as a reference. Results: We found that the relative absorbed dose sensitivity was highly dependent on both readout mode and integration time of the OSL signal. For CW-OSL signals containing only the blue emission band, S{sub rel} was between 0.85 and 0.94 for 1more » s readouts and between 0.82 and 0.93 for 10 s readouts. Similarly, for P-OSL readouts containing only the blue emission band S{sub rel} ranged from 0.86 to 0.91, and 0.82 to 0.93 for 1 s and 10 s readouts, respectively. For OSLD signals containing only the UV emission band, S{sub rel} ranged from 1.00 to 1.46, and 0.97 to 1.30 for P-OSL readouts of 1 s and 10 s, respectively. Conclusion: For measurements of absorbed dose using Al{sub 2}O{sub 3}:C OSLDs in therapeutic proton beams, dependence on beam quality was smallest for readout protocols that selected the blue emission band with small integration times. DA Granville received financial support from the Natural Sciences and Engineering Research Council of Canada.« less
Authors:
 [1] ; ;  [2]
  1. Carleton Laboratory for Radiotherapy Physics, Carleton University, Ottawa, ON (Canada)
  2. The University of Texas MD Anderson Cancer Center, Houston, TX (United States)
Publication Date:
OSTI Identifier:
22496305
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 62 RADIOLOGY AND NUCLEAR MEDICINE; ABSORBED RADIATION DOSES; ALUMINIUM OXIDES; DOSEMETERS; DOSIMETRY; GY RANGE 01-10; KEV RANGE 01-10; PHOTON BEAMS; PROTON BEAMS; RADIATION DETECTORS; RADIOTHERAPY; READOUT SYSTEMS; SIGNALS