Radiation assessment to paediatric with F-18-FDG undergo whole-body PET/CT examination
- Kluster Oncology Science and Radiology, Advanced Medical Dental Institute, Universiti Sains Malaysia (USM), Bertam, Penang (Malaysia)
- Department of Nuclear Medicine, National Cancer Institute, No 4 Jalan P7, Presint 7, Putrajaya (Malaysia)
This study was carried out on wholebody radiation dose assessment to paediatrics patient who undergo PET/CT scanner at Institut Kanser Negara. Consist of 68 patients with varies of malignancies and epilepsy disease case covering age between 2 years to 12 years old. This is a retrospective study from 2010-2014. The use of PET/CT scanner as an advanced tool has been proven to give an extra radiation dose to the patient. It is because of the radiation exposure from the combination of both CT and PET scans rather than a single CT or PET scan. Furthermore, a study on radiation dose to paediatric patient undergoing PET/CT is rare in Malaysia. So, the aim of this study is to estimate the wholebody effective dose to paediatric patient in Malaysia. Effective dose from PET scan was calculated based on the activity of F18 FDG and dose coefficient reported in International Commission on Radiological Protection (ICRP) Publication 106. Effective dose from CT was determined using k coefficient as reported in ICRP publication 102 and Dose Length Product (DLP) value. The average effective dose from PET and CT were found to be 7.05mSv and 5.77mSv respectively. The mean wholebody effective dose received by a patient with combined PETCT examination was 12.78mSv. These results could be used as reference for dosimetry of a patient undergoing PETCT examination in Malaysia.
- OSTI ID:
- 22494528
- Journal Information:
- AIP Conference Proceedings, Vol. 1704, Issue 1; Conference: iNuSTEC2015: International muclear science, technology and engineering conference 2015, Negeri Sembilan (Malaysia), 17-19 Aug 2015; Other Information: (c) 2016 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Patient-specific radiation dose and cancer risk estimation in CT: Part II. Application to patients
Comparison of FDG-PET/CT and CT for Delineation of Lumpectomy Cavity for Partial Breast Irradiation