skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-J-38: Improved DRR Image Quality Using Polyetheretherketone (PEEK) Fiducial in Image Guided Radiotherapy (IGRT)

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924125· OSTI ID:22494064
; ; ; ; ; ;  [1]
  1. Univ Alabama Birmingham, Birmingham, AL (United States)

Purpose Fiducial-based imaging is often used in IGRT. Traditional gold fiducial marker often has substantial reconstruction artifacts. These artifacts Result in poor image quality of DRR for online kV-to-DRR matching. This study evaluated the image quality of PEEK in DRR in static and moving phantom. Methods CT scan of the Gold and PEEK fiducial (both 1×3 mm) was acquired in a 22 cm cylindrical phantom filled with water. Image artifacts was evaluated with maximum CT value deviated from water due to artifacts; volume of artifacts in 10×10 cm in the center slice; maximum length of streak artifacts from the fiducial. DRR resolution were measured using FWHM and FWTM. 4DCT of PEEK fiducial was acquired with the phantom moving sinusoidally in superior-inferior direction. Motion artifacts were assessed for various 4D phase angles. Results The maximum CT value deviation was −174 for Gold and −24 for PEEK. The volume of artifacts in a 10x10 cm 3 mm slice was 0.369 for Gold and 0.074 cm3 for PEEK. The maximum length of streak artifact was 80mm for Gold and 7 mm for PEEK. PEEK in DRR, FWHM was close to actual (1.0 mm for Gold and 1.1 mm for PEEK). FWTM was 1.8 mm for Gold and 1.3 mm for PEEK in DRR. Barrel motion artifact of PEEK fiducial was noticeable for free-breathing scan. The apparent PEEK length due to residual motion was in close agreement with the calculated length (13 mm for 30–70 phase, 10 mm in 40–60 phase). Conclusion Streak artifacts on planning CT associated with use of gold fiducial can be significantly reduced by PEEK fiducial, while having adequate kV image contrast. DRR image resolution at FWTM was improved from 1.8 mm to 1.3 mm. Because of this improvement, we have been routinely use PEEK for liver IGRT.

OSTI ID:
22494064
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English