skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-J-33: Comparison Between Soft Tissue Alignment and Bony Alignment for Pancreatic Cancer Radiotherapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924120· OSTI ID:22494059
; ; ; ; ;  [1]
  1. UT MD Anderson Cancer Center, Houston, TX (United States)

Purpose An IGRT modality for pancreatic cancer treatment with dose escalation at our institution is in-room daily CT imaging. The purpose of this study is to assess the difference between soft tissue alignment and bony alignment for pancreatic tumor localization. Methods Eighteen patients with pancreatic tumors who underwent IMRT treatment with an inspiration breath-hold technique between July 2012 and February 2015 are included in this study. Prior to each treatment, a CT scan was acquired. The CT image guidance started with auto-alignment to either the bony anatomy (vertebral bodies) or fiducials (for the six patients with the stent in/near the tumor) and then, when necessary, manual adjustments were made based on soft tissue alignment using clinical software (CT-Assisted Targeting system). The difference between soft tissue alignment and bony/fiducial alignment was evaluated. Results Of all 380 treatments, manual adjustment was made in 225 treatments, ranging from 11% (3 treatments out of 28) to 96% (27 treatments out of 28) per patient. The mean of the difference between soft tissue alignment and bony/fiducial alignment per patient ranged from −3.6 to 0.3 mm, −1.5 to 2.8 mm, and −3.3 to 3.4 mm in the AP, SI, and RL directions, respectively. The maximum difference over all treatments was −9.5, −14.6, and −14.6 mm in the AP, SI, and RL directions, respectively. Conclusion About 60% of the time, manual adjustment based on soft tissue alignment was required. The extent of manual adjustment was usually small but varied significantly from patient to patient. The ultimate goal of the IGRT modality using daily CT imaging is not to fully cover the target but to spare organs-at-risk as much as possible to avoid them moving into higher dose gradients than accepted in the treatment plan. To this end, manual adjustment based on soft tissue alignment is critically important.

OSTI ID:
22494059
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English