Title: SU-E-I-26: The CT Compatibility of a Novel Direction Modulated Brachytherapy (DMBT) Tandem Applicator for Cervical Cancer

Abstract

Purpose: To examine CT metal image artifact from a novel direction-modulated brachytherapy (DMBT) tandem applicator (95% tungsten) for cervical cancer using a commercially available orthopedic metal artifact reduction (O-MAR) algorithm. Comparison to a conventional stainless steel applicator is also performed. Methods: Each applicator was placed in a water-filled phantom resembling the female pelvis and scanned in a Philips Brilliance 16-slice CT scanner using two pelvis protocols: a typical clinical protocol (120kVp, 16×0.75mm collimation, 0.692 pitch, 1.0s rotation, 350mm field of view (FOV), 600mAs, 1.5mm slices) and a protocol with a higher kVp and mAs setting useful for larger patients (140kVp, 16×0.75mm collimation, 0.688 pitch, 1.5s rotation, 350mm FOV, 870mAs, 1.5mm slices). Images of each tandem were acquired with and without the application of the O-MAR algorithm. Baseline scans of the phantom (no applicator) were also collected. CT numbers were quantified at distances from 5 to 30 mm away from the applicator’s edge (in increments of 5mm) using measurements at eight angles around the applicator, on three consecutive slices. Results: While the presence of both applicators degraded image quality, the DMBT applicator resulted in larger streaking artifacts and dark areas in the image compared to the stainless steel applicator. Application ofmore » the O-MAR algorithm improved all acquired images, both visually and quantitatively. The use of low and high kVp and mAs settings (120 kVp/600mAs and 140 kVp/870mAs) in conjunction with the O-MAR algorithm lead to similar CT numbers in the vicinity of the applicator and a similar reduction of the induced metal artifact. Conclusion: This work indicated that metal artifacts induced by the DMBT and the stainless steel applicator are greatly reduced when using the O-MAR algorithm, leading to better quality phantom images. The use of a high dose protocol provided similar improvements in metal artifacts compared to the clinical protocol.« less

Authors:
; ; ; ;  [1];  [2];  [1]
  1. Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario (Canada)
  2. The Netherlands Cancer Institute, Amsterdam (Netherlands)
Publication Date:
OSTI Identifier:
22493985
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; ALGORITHMS; BRACHYTHERAPY; COMPUTERIZED TOMOGRAPHY; IMAGES; NEOPLASMS; PATIENTS; PELVIS; PHANTOMS; RADIATION DOSES; STAINLESS STEELS; TUNGSTEN

Citation Formats

Elzibak, A, Safigholi, H, Soliman, A, Ravi, A, Song, WY, Kager, P, Han, D, and University of California, San Diego, La Jolla, CA. SU-E-I-26: The CT Compatibility of a Novel Direction Modulated Brachytherapy (DMBT) Tandem Applicator for Cervical Cancer. United States: N. p., 2015. Web. doi:10.1118/1.4924023.
Elzibak, A, Safigholi, H, Soliman, A, Ravi, A, Song, WY, Kager, P, Han, D, & University of California, San Diego, La Jolla, CA. SU-E-I-26: The CT Compatibility of a Novel Direction Modulated Brachytherapy (DMBT) Tandem Applicator for Cervical Cancer. United States. https://doi.org/10.1118/1.4924023
Elzibak, A, Safigholi, H, Soliman, A, Ravi, A, Song, WY, Kager, P, Han, D, and University of California, San Diego, La Jolla, CA. 2015. "SU-E-I-26: The CT Compatibility of a Novel Direction Modulated Brachytherapy (DMBT) Tandem Applicator for Cervical Cancer". United States. https://doi.org/10.1118/1.4924023.
@article{osti_22493985,
title = {SU-E-I-26: The CT Compatibility of a Novel Direction Modulated Brachytherapy (DMBT) Tandem Applicator for Cervical Cancer},
author = {Elzibak, A and Safigholi, H and Soliman, A and Ravi, A and Song, WY and Kager, P and Han, D and University of California, San Diego, La Jolla, CA},
abstractNote = {Purpose: To examine CT metal image artifact from a novel direction-modulated brachytherapy (DMBT) tandem applicator (95% tungsten) for cervical cancer using a commercially available orthopedic metal artifact reduction (O-MAR) algorithm. Comparison to a conventional stainless steel applicator is also performed. Methods: Each applicator was placed in a water-filled phantom resembling the female pelvis and scanned in a Philips Brilliance 16-slice CT scanner using two pelvis protocols: a typical clinical protocol (120kVp, 16×0.75mm collimation, 0.692 pitch, 1.0s rotation, 350mm field of view (FOV), 600mAs, 1.5mm slices) and a protocol with a higher kVp and mAs setting useful for larger patients (140kVp, 16×0.75mm collimation, 0.688 pitch, 1.5s rotation, 350mm FOV, 870mAs, 1.5mm slices). Images of each tandem were acquired with and without the application of the O-MAR algorithm. Baseline scans of the phantom (no applicator) were also collected. CT numbers were quantified at distances from 5 to 30 mm away from the applicator’s edge (in increments of 5mm) using measurements at eight angles around the applicator, on three consecutive slices. Results: While the presence of both applicators degraded image quality, the DMBT applicator resulted in larger streaking artifacts and dark areas in the image compared to the stainless steel applicator. Application of the O-MAR algorithm improved all acquired images, both visually and quantitatively. The use of low and high kVp and mAs settings (120 kVp/600mAs and 140 kVp/870mAs) in conjunction with the O-MAR algorithm lead to similar CT numbers in the vicinity of the applicator and a similar reduction of the induced metal artifact. Conclusion: This work indicated that metal artifacts induced by the DMBT and the stainless steel applicator are greatly reduced when using the O-MAR algorithm, leading to better quality phantom images. The use of a high dose protocol provided similar improvements in metal artifacts compared to the clinical protocol.},
doi = {10.1118/1.4924023},
url = {https://www.osti.gov/biblio/22493985}, journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 42,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2015},
month = {Mon Jun 15 00:00:00 EDT 2015}
}