Title: Polyhydroxybutyrate/valerate/polycaprolactone small-diameter vascular graft: Experimental study of integration into organism

We prepared polyhydroxybutyrate/valerate (PHBV)/polylcaprolactone (PCL) small-diameter vascular grafts using electrospinning. Surface structure was assessed by scanning electron microscopy whilst physicomechanical properties were investigated by longitudinal uniaxial tension testing. Patency of grafts implanted into the rat abdominal aorta was evaluated using a Doppler ultrasonography at 2 week, 1 month and 12 month postimplantation. In addition, we assessed local histological features, along with IL-1β, IL-2, IL-4, IL-10, TNFa, TGF-β1, and VEGF serum levels. We revealed that only 2 (25%) grafts were not thrombosed at 2 week and 1 month postimplantation. However, at 12 month postimplantation a satisfactory histological pattern was observed in 50% of all cases, and we detected a monolayer of endothelial cells on the inner graft surface in half the cases. Regarding other grafts, we revealed minor connective tissue hyperplasia in 41.7% of the grafts and an inflammatory infiltrate in the part of the arterial wall in 8.3% of the grafts. We found that the IL-1β serum level was 3.5-fold higher in the group of experimental rats at 12 month postimplantation (p < 0.01). In addition, the IL-2 and IL-4 serum levels at 12 month postimplantation were 2- and 2.8-fold higher as compared to short-term implantation (2 weeks and 1more » month) and control rats (p < 0.05) whilst the IL-10 serum level at 1 and 12 month postimplantation was more than 100-fold higher in comparison with 2 week postimplantation and control rats (p < 0.001). Serum VEGF was detected only at 12 month postimplantation. All in all, we created a biocompatible PHBV/PCL small-diameter vascular graft with a high surface area to volume ratio. A long-term patency of biodegradable vascular grafts after implantation into the rat abdominal aorta and the absence of a considerable immune response confirmed a high biocompatibility of such construct and the possibility of its use as a vascular graft.« less
Authors:
; ; ; ; ; ; ; ;  [1] ;  [2]
  1. Research Institute for Complex Issues of Cardiovascular Diseases, Kemerovo (Russian Federation)
  2. Kemerovo Cardiology Dispensary, Kemerovo (Russian Federation)
Publication Date:
OSTI Identifier:
22492508
Resource Type:
Journal Article
Resource Relation:
Journal Name: AIP Conference Proceedings; Journal Volume: 1683; Journal Issue: 1; Conference: International conference on advanced materials with hierarchical structure for new technologies and reliable structures 2015, Tomsk (Russian Federation), 21-25 Sep 2015; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; AORTA; BIODEGRADATION; COMPARATIVE EVALUATIONS; CONNECTIVE TISSUE; DOPPLER EFFECT; GRAFTS; INFLAMMATION; MECHANICAL TESTS; RATS; SCANNING ELECTRON MICROSCOPY; SURFACE AREA; SURFACE TENSION; SURFACES; ULTRASONOGRAPHY