Title: SU-C-304-02: Robust and Efficient Process for Acceptance Testing of Varian TrueBeam Linacs Using An Electronic Portal Imaging Device (EPID)

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4923790· OSTI ID:22486537
; ; ; ; ; ;  [1];  [2];  [3]
  1. Washington University School of Medicine, Saint Louis, MO (United States)
  2. Oregon State University, Corvallis, OR (United States)
  3. UCSD Medical Center, La Jolla, CA (United States)

Purpose: The purpose of this project was to develop a process that utilizes the onboard kV and MV electronic portal imaging devices (EPIDs) to perform rapid acceptance testing (AT) of linacs in order to improve efficiency and standardize AT equipment and processes. Methods: In this study a Varian TrueBeam linac equipped with an amorphous silicon based EPID (aSi1000) was used. The conventional set of AT tests and tolerances was used as a baseline guide, and a novel methodology was developed to perform as many tests as possible using EPID exclusively. The developer mode on Varian TrueBeam linac was used to automate the process. In the current AT process there are about 45 tests that call for customer demos. Many of the geometric tests such as jaw alignment and MLC positioning are performed with highly manual methods, such as using graph paper. The goal of the new methodology was to achieve quantitative testing while reducing variability in data acquisition, analysis and interpretation of the results. The developed process was validated on two machines at two different institutions. Results: At least 25 of the 45 (56%) tests which required customer demo can be streamlined and performed using EPIDs. More than half of the AT tests can be fully automated using the developer mode, while others still require some user interaction. Overall, the preliminary data shows that EPID-based linac AT can be performed in less than a day, compared to 2–3 days using conventional methods. Conclusions: Our preliminary results show that performance of onboard imagers is quite suitable for both geometric and dosimetric testing of TrueBeam systems. A standardized AT process can tremendously improve efficiency, and minimize the variability related to third party quality assurance (QA) equipment and the available onsite expertise. Research funding provided by Varian Medical Systems. Dr. Sasa Mutic receives compensation for providing patient safety training services from Varian Medical Systems, the sponsor of this study.

OSTI ID:
22486537
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English