Title: Heating system of pellet samples integrated with terahertz spectrometer

This article describes automation of temperature-dependent terahertz spectroscopic experiments. The proposed dual-heater temperature controller based on a cascade proportional-integral-derivative algorithm provides smooth temperature changes in the polyethylene-based pharmaceutical pellet samples. The device has been integrated with a terahertz time-domain spectrometer. Thermodynamic experiments can now be performed without any probe inserted into the measured sample. Selected results of temperature-induced evolution in terahertz spectra are presented.
Authors:
; ;  [1]
  1. Department of Electronics, Wroclaw University of Technology, 27 Wybrzeze Wyspianskiego St., 50-370 Wroclaw (Poland)
Publication Date:
OSTI Identifier:
22482811
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 87; Journal Issue: 1; Other Information: (c) 2016 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ALGORITHMS; AUTOMATION; HEATING SYSTEMS; PELLETS; POLYETHYLENES; SPECTRA; SPECTROMETERS; TEMPERATURE DEPENDENCE