skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: A laboratory 8 keV transmission full-field x-ray microscope with a polycapillary as condenser for bright and dark field imaging

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4929602· OSTI ID:22482746
 [1]; ;  [2];  [3]
  1. Institute for X-Optics, University of Applied Sciences Koblenz, RheinAhrCampus Remagen, Joseph-Rovan-Allee 2, D-53424 Remagen (Germany)
  2. Institute for Optics and Atomic Physics, Technical University of Berlin, Hardenbergstrasse 36, D-10623 Berlin (Germany)
  3. Max-Born-Institute, Max-Born-Strasse 2A, D-12489 Berlin (Germany)

This article introduces a laboratory setup of a transmission full-field x-ray microscope at 8 keV photon energy. The microscope operates in bright and dark field imaging mode with a maximum field of view of 50 μm. Since the illumination geometry determines whether the sample is illuminated homogeneously and moreover, if different imaging methods can be applied, the condenser optic is one of the most significant parts. With a new type of x-ray condenser, a polycapillary optic, we realized bright field imaging and for the first time dark field imaging at 8 keV photon energy in a laboratory setup. A detector limited spatial resolution of 210 nm is measured on x-ray images of Siemens star test patterns.

OSTI ID:
22482746
Journal Information:
Review of Scientific Instruments, Vol. 86, Issue 8; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English