Title: Note: A large open ratio, time, and position sensitive detector for time of flight measurements in UHV

We report on the construction of an UHV compatible 40 mm active diameter detector based on micro channel plates and assembled directly on the feed-throughs of a DN63CF flange. It is based on the charge division technique and uses a standard 2 inch Si wafer as a collector. The front end electronic is placed directly on the air side of the flange allowing excellent immunity to noise and a very good timing signal with reduced ringing. The important aberrations are corrected empirically providing an absolute positioning accuracy of 500 μm while a 150 μm resolution is measured in the center.
Authors:
; ; ; ; ; ;  [1]
  1. Institut des Sciences Moléculaires d’Orsay (ISMO), CNRS, Univ. Paris-Sud, Université Paris-Saclay, Orsay F-91405 (France)
Publication Date:
OSTI Identifier:
22482686
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 86; Journal Issue: 12; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ACCURACY; CONSTRUCTION; FLANGES; MICROCHANNEL ELECTRON MULTIPLIERS; NOISE; POSITION SENSITIVE DETECTORS; POSITIONING; RESOLUTION; SIGNALS; TIME-OF-FLIGHT METHOD