Title: Verification and compensation of respiratory motion using an ultrasound imaging system

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4907958· OSTI ID:22413464
; ; ;  [1]; ;  [2]
  1. Department of Mechanical Engineering, National Taipei University of Technology, Number 1, Section 3, Chung-Hsiao E. Road, Taipei 10608, Taiwan (China)
  2. Shin Kong Wu Ho-Su Memorial Hospital, Number 95, Wen-Chang Road, Shih-Lin District, Taipei 11101, Taiwan (China)

Purpose: The purpose of this study was to determine if it is feasible to use ultrasound imaging as an aid for moving the treatment couch during diagnosis and treatment procedures associated with radiation therapy, in order to offset organ displacement caused by respiratory motion. A noninvasive ultrasound system was used to replace the C-arm device during diagnosis and treatment with the aims of reducing the x-ray radiation dose on the human body while simultaneously being able to monitor organ displacements. Methods: This study used a proposed respiratory compensating system combined with an ultrasound imaging system to monitor the compensation effect of respiratory motion. The accuracy of the compensation effect was verified by fluoroscopy, which means that fluoroscopy could be replaced so as to reduce unnecessary radiation dose on patients. A respiratory simulation system was used to simulate the respiratory motion of the human abdomen and a strain gauge (respiratory signal acquisition device) was used to capture the simulated respiratory signals. The target displacements could be detected by an ultrasound probe and used as a reference for adjusting the gain value of the respiratory signal used by the respiratory compensating system. This ensured that the amplitude of the respiratory compensation signal was a faithful representation of the target displacement. Results: The results show that performing respiratory compensation with the assistance of the ultrasound images reduced the compensation error of the respiratory compensating system to 0.81–2.92 mm, both for sine-wave input signals with amplitudes of 5, 10, and 15 mm, and human respiratory signals; this represented compensation of the respiratory motion by up to 92.48%. In addition, the respiratory signals of 10 patients were captured in clinical trials, while their diaphragm displacements were observed simultaneously using ultrasound. Using the respiratory compensating system to offset, the diaphragm displacement resulted in compensation rates of 60%–84.4%. Conclusions: This study has shown that a respiratory compensating system combined with noninvasive ultrasound can provide real-time compensation of the respiratory motion of patients.

OSTI ID:
22413464
Journal Information:
Medical Physics, Vol. 42, Issue 3; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Reproducibility of liver position using active breathing coordinator for liver cancer radiotherapy
Journal Article · Wed Mar 01 00:00:00 EST 2006 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22413464

SU-C-209-02: 3D Fluoroscopic Image Generation From Patient-Specific 4DCBCT-Based Motion Models Derived From Clinical Patient Images
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22413464

A new method for tracking organ motion on diagnostic ultrasound images
Journal Article · Mon Sep 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22413464