Title: SU-C-19A-06: A Robust and Affordable Table Indexing Approach for Total Lymphoid Irradiation Treatment

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889705· OSTI ID:22412430
; ; ;  [1]
  1. Stanford University, Stanford, CA (United States)

Purpose: Total lymphoid irradiation (TLI) is conventionally delivered through the dosimetric matching of mantle, spleen, and pelvis fields, necessitating multiple isocenters delivered through a combination of couch shifts and sliding of patients relative to the couch rendering the technique susceptible to shifting errors. To address this challenge, a novel technique for the couch indexing of TLI treatments is developed and evaluated through a multi-patient pilot trial. Methods: An immobilization device was designed consisting of a movable indexed slide board with an Exact Lok-Bar drilled into it. A Timo headrests were used fixate the head of the patient relative to the slide board. For the Varian Exact Couch™, the immobilization board was connected to the H3 notch to avoid the metal infrastructure of the couch for the delivery of the mantle and spleen fields. For tall patients the required shift for the pelvis isocenter reaches the shifting limit and the board was slid from H3 to H4 (a fixed distance of 14 cm). A total 22 patients were stratified in two groups of 11, one consisting of the conventional setup, and one group with the proposed immobilization technique. Results: The standard deviations (SD) of the couch positions in lateral, longitudinal, and vertical directions for 10 fractions for each patient in both groups were calculated. In the non-indexed group, the positioning SD ranged from 0.9 to 4.7 cm. Using our device, the positioning SD was reduced to a range of 0.2 to 0.9 cm, with the longitudinal direction showing the largest improvement. Conclusion: Matched field TLI remains error prone to geometrical misses. The feasibility of full indexing TLI treatments was validated and shown to result in a significant reduction of positioning errors.

OSTI ID:
22412430
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Clinical Evaluation of a Robotic 6-Degree of Freedom Treatment Couch for Frameless Radiosurgery
Journal Article · Tue May 01 00:00:00 EDT 2012 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22412430

SU-E-T-248: Near Real-Time Analysis of Radiation Delivery and Imaging, Accuracy to Ensure Patient Safety
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22412430

SU-E-T-315: Dosimetric Effects of Couch Top Shift On VMAT Delivery in Absence of Indexing
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22412430