skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-C-16A-03: Direction Modulated Brachytherapy for HDR Treatment of Cervical Cancer

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889695· OSTI ID:22412421
; ; ; ; ; ;  [1];  [2];  [3]
  1. University of California, San Diego, La Jolla, CA (United States)
  2. McGill University, Montreal, QC (Canada)
  3. Sunnybrook Odette Cancer Centre, Toronto (Canada)

Purpose: To investigate a new Directional Modulated Brachytherapy (DMBT) intra-uterine tandem using various 192-Ir after-loaders. Methods: Dose distributions from the 192-Ir sources were modulated using a 6.3mm diameter tungsten shield (18.0g/cm3). The source moved along 6 longitudinal grooves, each 1.3mm in diameter, evenly spaced along periphery of the shield, The tungsten rod was enclosqed by 0.5mm thick Delrin (1.41g/cc). Monte Carlo N particle (MCNPX) was used to calculate dose distributions. 51million particles were calculated on 504 cores of a supercomputer. Fifteen different patients originally treated with a traditional tandem-and-ovoid applicator, with 5 fractions each, (15 patients X 5 fxs = 75 plans) were re-planned with the DMBT applicator combined with traditional ovoids, on an in-house developed HDR brachytherapy planning platform, which used intensity modulated planning capabilities using a constrained gradient optimization algorithm. For all plans the prescription dose was 6 Gy and they were normalized to match the clinical treated V100. Results: Generally, the DMBT plan quality was a remarkable improvement from conventional T and O plans because of the anisotropic dose distribution of DMBT. The largest difference was to the bladder which had a 0.59±0.87 Gy (8.5±28.7%) reduction in dose. This was because of the the horseshoe shape (U-shape) of the bladder. The dose reduction to rectum and sigmoid were 0.48±0.55 Gy (21.1±27.2%) and 0.10±0.38 Gy (40.6±214.9%), respectively. The D90 to the HRCTV was 6.55±0.96 Gy (conventional T and O) and 6.59±1.06 Gy (DMBT). Conclusion: For image guided adaptive brachytherapy, greater flexibility of radiation intensity is essential and DMBT can be the solution.

OSTI ID:
22412421
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English