Title: TH-A-18C-11: An Investigation of KV CBCT Image Quality and Dose Reduction for Volume-Of-Interest Imaging Using Dynamic Collimation

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889570· OSTI ID:22409813
 [1];  [1]
  1. of Physics and Atmospheric Science, Dalhousie University, Halifax, NS (Canada)

Purpose: The focus of this work was to investigate the improvements in image quality and dose reduction for volume-of-interest (VOI) kV-CBCT using dynamic collimation. Methods: A prototype iris aperture was used to track a VOI during a CBCT acquisition. The current aperture design is capable of one-dimensional translation as a function of gantry angle and dynamic adjustment of the iris radius. The aperture occupies the location of the bow-tie filter on a Varian OBI system. CBCT and planar image quality was investigated as a function of aperture radius, while maintaining the same dose to the VOI, for a 20 cm diameter cylindrical water phantom with a 9 mm diameter bone insert centered on isocenter. Corresponding scatter-to-primary ratios (SPR) were determined at the detector plane with Monte Carlo simulation using EGSnrc. Dose distributions for various anatomical sites were modeled using a dynamic BEAMnrc library and DOSXYZnrc. The resulting VOI dose distributions were compared to full-field distributions. Results: SPR was reduced by a factor of 8.4 when decreasing iris diameter from 21.2 cm to 2.4 cm (at isocenter). Similarly, this change in iris diameter corresponds to a factor increase of approximately 1.4 and 1.5 in image contrast for CBCT and planar images, respectively, and similarly a factor decrease in image noise of approximately 1.7 and 1.5. This results in a measured gain in contrast-to-noise ratio of a factor of approximately 2.3 for both CBCT and planar images. Depending upon the anatomical site, dose was reduced to 10%–70% of the full field value along the central axis plane and down to 2% along the axial planes, while maintaining the same dose to the VOI compared to full-field techniques. Conclusion: The presented VOI technique offers improved image quality for image-guided radiotherapy while sparing the surrounding volume of unnecessary dose compared to full-field techniques.

OSTI ID:
22409813
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English