Title: TU-C-BRE-03: Aggregation of Linac Measurement Data

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889266· OSTI ID:22409664
; ; ; ; ; ;  [1]
  1. UT MD Anderson Cancer Center, Houston, TX (United States)

Purpose: Accurate data of linear accelerator radiation characteristics is important for treatment planning system commissioning as well as regular quality assurance of the machine. The RPC has performed site visits of numerous machines . Data gathered from Varian machines from the past 15 years are presented. The data collected can be used as a secondary check or when commissioning a new machine to verify that values are reasonable. Methods: Data from the past 15 years of RPC site visits was compiled and analyzed. Data was composed from measurements from approximately 400 Varian machines. Each dataset consists of several point measurements at various locations in a water phantom to measure percentage depth dose, output factors, including small MLC fields, off-axis factors, and wedge factors if applicable. Common statistical values are presented for each machine type. Where applicable, data was compared to other reference data given by the vendor or a select number of previous researchers. Results: Data is separated by energy and parameter and then analyzed by machine class. Data distributions of the parameter data were normal except occasionally at the tails. Distributions of the data for each class and parameter are tabulated to give not simply a singular reference value, but metrics about the distribution: 5th and 95th percentile values and the standard deviation as well as the median. Conclusion: The RPC has collected numerous data on Varian linacs and presented the finding of the past 15 years. The data can be used as a reference data set for physicists to compare against. A linac that deviates from the values does not necessarily indicate there is a problem as long as the treatment planning system correlates to the machine. Comparison of linac and treatment planning system data to external reference data can prevent serious treatment errors.

OSTI ID:
22409664
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English