Title: Reduction of ring artifacts in CBCT: Detection and correction of pixel gain variations in flat panel detectors

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4893278· OSTI ID:22409550
 [1]; ; ;  [2]
  1. Department of Radiation Oncology, University of Colorado School of Medicine, Aurora, Colorado 80045 (United States)
  2. Department of Imaging Physics, University of Texas M.D. Anderson Cancer Center, Houston, Texas 77030 (United States)

Purpose: In using flat panel detectors (FPD) for cone beam computed tomography (CBCT), pixel gain variations may lead to structured nonuniformities in projections and ring artifacts in CBCT images. Such gain variations can be caused by change in detector entrance exposure levels or beam hardening, and they are not accounted by conventional flat field correction methods. In this work, the authors presented a method to identify isolated pixel clusters that exhibit gain variations and proposed a pixel gain correction (PGC) method to suppress both beam hardening and exposure level dependent gain variations. Methods: To modulate both beam spectrum and entrance exposure, flood field FPD projections were acquired using beam filters with varying thicknesses. “Ideal” pixel values were estimated by performing polynomial fits in both raw and flat field corrected projections. Residuals were calculated by taking the difference between measured and ideal pixel values to identify clustered image and FPD artifacts in flat field corrected and raw images, respectively. To correct clustered image artifacts, the ratio of ideal to measured pixel values in filtered images were utilized as pixel-specific gain correction factors, referred as PGC method, and they were tabulated as a function of pixel value in a look-up table. Results: 0.035% of detector pixels lead to clustered image artifacts in flat field corrected projections, where 80% of these pixels were traced back and linked to artifacts in the FPD. The performance of PGC method was tested in variety of imaging conditions and phantoms. The PGC method reduced clustered image artifacts and fixed pattern noise in projections, and ring artifacts in CBCT images. Conclusions: Clustered projection image artifacts that lead to ring artifacts in CBCT can be better identified with our artifact detection approach. When compared to the conventional flat field correction method, the proposed PGC method enables characterization of nonlinear pixel gain variations as a function of change in x-ray spectrum and intensity. Hence, it can better suppress image artifacts due to beam hardening as well as artifacts that arise from detector entrance exposure variation.

OSTI ID:
22409550
Journal Information:
Medical Physics, Vol. 41, Issue 9; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

WE-E-18A-07: MAGIC: Multi-Acquisition Gain Image Correction for Mobile X-Ray Systems with Intrinsic Localization Crosshairs
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22409550

A model-based scatter artifacts correction for cone beam CT
Journal Article · Fri Apr 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22409550

Shading correction for on-board cone-beam CT in radiation therapy using planning MDCT images
Journal Article · Fri Oct 15 00:00:00 EDT 2010 · Medical Physics · OSTI ID:22409550