Title: Poster — Thur Eve — 15: Improvements in the stability of the tomotherapy imaging beam

Use of helical TomoTherapy based MVCT imaging for adaptive planning requires the image values (HU) to remain stable over the course of treatment. In the past, the image value stability was suboptimal, which required frequent change to the image value to density calibration curve to avoid dose errors on the order of 2–4%. The stability of the image values at our center was recently improved by stabilizing the dose rate of the machine (dose control servo) and performing daily MVCT calibration corrections. In this work, we quantify the stability of the image values over treatment time by comparing patient treatment image density derived using MVCT and KVCT. The analysis includes 1) MVCT - KVCT density difference histogram, 2) MVCT vs KVCT density spectrum, 3) multiple average profile density comparison and 4) density difference in homogeneous locations. Over two months, the imaging beam stability was compromised several times due to a combination of target wobbling, spectral calibration, target change and magnetron issues. The stability of the image values were analyzed over the same period. Results show that the impact on the patient dose calculation is 0.7% +− 0.6%.
Authors:
 [1]
  1. The Ottawa Hospital Cancer Center, Ontario (Canada)
Publication Date:
OSTI Identifier:
22409529
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 8; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; CALIBRATION; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; DOSE RATES; MAGNETRONS; PATIENTS; PLANNING; RADIATION DOSES