Title: Kilovoltage cone-beam CT imaging dose during breast radiotherapy: A dose comparison between a left and right breast setup

The purpose of this study was to investigate the delivered dose from a kilovoltage cone-beam computed tomography (kV-CBCT) acquired in breast treatment position for a left and right breast setup. The dose was measured with thermoluminescent dosimeters positioned within a female anthropomorphic phantom at organ locations. Imaging was performed on an Elekta Synergy XVI system with the phantom setup on a breast board. The image protocol involved 120 kVp, 140 mAs, and a 270° arc rotation clockwise 0° to 270° for the left breast setup and 270° to 180° for the right breast setup (maximum arc rotations possible). The dose delivered to the left breast, right breast, and heart was 5.1 mGy, 3.9 mGy, and 4.0 mGy for the left breast setup kV-CBCT, and 6.4 mGy, 6.0 mGy, and 4.8 mGy for the right breast setup kV-CBCT, respectively. The rotation arc of the kV-CBCT influenced the dose delivered, with the right breast setup kV-CBCT found to deliver a dose of up to 4 mGy or 105% higher to the treated breast′s surface in comparison with the left breast setup. This is attributed to the kV-CBCT source being more proximal to the anterior of the phantom for a right breast setup,more » whereas the source is more proximal to the posterior of the patient for a left-side scan.« less
Authors:
 [1] ;  [2] ;  [2] ;  [3] ;  [2] ;  [2] ;  [2] ;  [2] ; ;  [4] ;  [3] ;  [2] ;  [2]
  1. Northern Sydney Cancer Centre, Royal North Shore Hospital, NSW (Australia)
  2. (Australia)
  3. Centre for Medical Radiation Physics, University of Wollongong, NSW (Australia)
  4. Liverpool and Macarthur Cancer Therapy Centres, NSW (Australia)
Publication Date:
OSTI Identifier:
22395917
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Dosimetry; Journal Volume: 39; Journal Issue: 2; Other Information: Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; COMPUTERIZED TOMOGRAPHY; IMAGES; MAMMARY GLANDS; PATIENTS; PHANTOMS; RADIATION DOSES; RADIATION PROTECTION; RADIOTHERAPY; THERMOLUMINESCENT DOSEMETERS