Title: Measuring skewness of red blood cell deformability distribution by laser ektacytometry

An algorithm is proposed for measuring the parameters of red blood cell deformability distribution based on laser diffractometry of red blood cells in shear flow (ektacytometry). The algorithm is tested on specially prepared samples of rat blood. In these experiments we succeeded in measuring the mean deformability, deformability variance and skewness of red blood cell deformability distribution with errors of 10%, 15% and 35%, respectively. (laser biophotonics)
Authors:
; ;  [1] ;  [2]
  1. International Laser Center, M. V. Lomonosov Moscow State University, Moscow (Russian Federation)
  2. M. V. Lomonosov Moscow State University, Faculty of Computational Mathematics and Cybernetics, Moscow (Russian Federation)
Publication Date:
OSTI Identifier:
22395829
Resource Type:
Journal Article
Resource Relation:
Journal Name: Quantum Electronics (Woodbury, N.Y.); Journal Volume: 44; Journal Issue: 8; Other Information: Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; 72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS; ALGORITHMS; ASYMMETRY; BLOOD CELLS; DIFFRACTION; DISTRIBUTION; ERRORS; LASERS; RATS; STATISTICS