skip to main content

Title: Fibre optics wavemeters calibration using a self-referenced optical frequency comb

Self-referenced optical frequency combs enable the measurement of optical frequencies with a very high accuracy, achieving uncertainties close to the atomic clock used as reference (<10{sup −13} s). In this paper, we present the technique for the measurement of laser frequencies for optical communications followed at IO-CSIC and its application to the calibration of two wavemeters in the 1.5 μm optical communication window. Calibration uncertainties down to 12 MHz and 59 MHz were obtained, respectively, for each of the devices. Furthermore, the long-term behaviour of the higher resolution wavemeter was studied during a 750 h period of sustained operation, exhibiting a dispersion in the measurements of 7.72 MHz. Temperature dependence of the device was analysed, enabling to further reduce dispersion down to a 2.15 MHz range, with no significant temporal deviations.
Authors:
; ;  [1]
  1. Instituto de Óptica IO-CSIC, C/Serrano 121, 28006 Madrid (Spain)
Publication Date:
OSTI Identifier:
22392304
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 86; Journal Issue: 1; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; ACCURACY; CALIBRATION; COMMUNICATIONS; DISPERSIONS; EQUIPMENT; FIBERS; LASERS; MHZ RANGE 01-100; OPERATION; OPTICS; RESOLUTION; TEMPERATURE DEPENDENCE