Title: Image plate characterization and absolute calibration to low kilo-electron-volt electrons

We report on the characterization of an image plate and its absolute calibration to electrons in the low keV energy range (1–30 keV). In our case, an Agfa MD4.0 without protection layer was used in combination with a Fuji FLA7000 scanner. The calibration data are compared to other published data and a consistent picture of the sensitivity of image plates to electrons is obtained, which suggests a validity of the obtained calibration up to 100 keV.
Authors:
 [1] ;  [2] ; ; ;  [3]
  1. GSI Helmholtzzentrum für Schwerionenforschung, Planckstraße 1, 64291 Darmstadt (Germany)
  2. (Germany)
  3. Institus für Kernphysik, Technische Universität Darmstadt, Schlossgartenstraße 9, 64289 Darmstadt (Germany)
Publication Date:
OSTI Identifier:
22392228
Resource Type:
Journal Article
Resource Relation:
Journal Name: Review of Scientific Instruments; Journal Volume: 85; Journal Issue: 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; CALIBRATION; COMPARATIVE EVALUATIONS; ELECTRONS; IMAGES; KEV RANGE 01-10; KEV RANGE 10-100; LAYERS; PLATES; SAFETY; SENSITIVITY