Title: Dosimetric study of surface applicators of HDR brachytherapy GammaMed Plus equipment

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4901388· OSTI ID:22390515

The cone type surface applicators used in HDR brachytherapy for treatment of small skin lesions are an alternative to be used with both electron beams and orthovoltage X-ray equipment. For a good treatment planning is necessary to know the dose distribution of these applicators, which can be obtained by experimental measurement and Monte Carlo simulation as well. In this study the dose distribution of surface applicators of 3 and 3.5 cm diameter, respectively of HDR brachytherapy GammaMed Plus equipment has been estimated using the Monte Carlo method, MCNP code. The applicators simulated were placed on the surface of a water phantom of 20 × 20 × 20 cm and the dose was calculated at depths from 0 to 3 cm with increments of 0.25 mm. The dose profiles obtained at depth show the expected gradients for surface therapy.

OSTI ID:
22390515
Journal Information:
AIP Conference Proceedings, Vol. 1626, Issue 1; Conference: 13. Mexican Symposium on Medical Physics, Leon, Guanajuato (Mexico), 15-16 Mar 2014; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English