Title: SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template

Abstract

Purpose: HDR brachytherapy using interstitial needle template for cervical cancer is commonly delivered in 4-5 fractions. Routine verification of needle positions before each fraction is often based on radiographic imaging of implanted fiducial markers. The current study evaluated interfractional displacement of implanted fiducial markers using CT images. Methods: 9 sequential patients with cervical interstitial needle implants were evaluated. The superior and inferior borders of the target volumes were defined by fiducial markers in planning CT. The implant position was verified with kV orthogonal images before each fraction. A second CT was acquired prior 3rd fraction (one or 2 days post planning CT). Distances from inferior and superior fiducial markers to pubic symphysis plane (perpendicular to vaginal obtulator)were measured. Distance from needle tip of a reference needle (next to the inferior marker) to the pubic symphysis plane was also determined. The difference in fiducial marker distance or needle tip distance between planning CT and CT prior 3rd fraction were measured to assess markers migration and needle displacement. Results: The mean inferior marker displacement was 4.5 mm and ranged 0.9 to 11.3 mm. The mean superior marker displacement was 2.7 mm and ranged 0 to 10.4 mm. There was a good associationmore » between inferior and superior marker displacement (r=0.95). Mean averaged inferior and superior marker displacement was 3.3 mm and ranged from 0.1 to 10.9 mm, with a standard deviation of 3.2 mm. The mean needle displacement was 5.6 mm and ranged 0.2 to 15.6 mm. Needle displacements were reduced (p<0.05) after adjusting according to needle-to-fiducials distance. Conclusion: There were small fiducial marker displacements between HDR fractions. Our study suggests a target margin of 9.7 mm to cover interfractional marker displacements (in 95% cases) for pretreatment verification based on radiographic imaging.« less

Authors:
; ; ; ; ; ; ; ; ;  [1]
  1. UniversityAlabama Birmingham, Birmingham, AL (United Kingdom)
Publication Date:
OSTI Identifier:
22369746
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 41; Journal Issue: 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; BRACHYTHERAPY; IMAGES; IMPLANTS; NEOPLASMS; PATIENTS; VERIFICATION

Citation Formats

Shen, S, Kim, R, Benhabib, S, Araujo, J, Burnett, L, Duan, J, Popple, R, Wu, X, Cardan, R, and Brezovich, I. SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template. United States: N. p., 2014. Web. doi:10.1118/1.4888840.
Shen, S, Kim, R, Benhabib, S, Araujo, J, Burnett, L, Duan, J, Popple, R, Wu, X, Cardan, R, & Brezovich, I. SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template. United States. https://doi.org/10.1118/1.4888840
Shen, S, Kim, R, Benhabib, S, Araujo, J, Burnett, L, Duan, J, Popple, R, Wu, X, Cardan, R, and Brezovich, I. 2014. "SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template". United States. https://doi.org/10.1118/1.4888840.
@article{osti_22369746,
title = {SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template},
author = {Shen, S and Kim, R and Benhabib, S and Araujo, J and Burnett, L and Duan, J and Popple, R and Wu, X and Cardan, R and Brezovich, I},
abstractNote = {Purpose: HDR brachytherapy using interstitial needle template for cervical cancer is commonly delivered in 4-5 fractions. Routine verification of needle positions before each fraction is often based on radiographic imaging of implanted fiducial markers. The current study evaluated interfractional displacement of implanted fiducial markers using CT images. Methods: 9 sequential patients with cervical interstitial needle implants were evaluated. The superior and inferior borders of the target volumes were defined by fiducial markers in planning CT. The implant position was verified with kV orthogonal images before each fraction. A second CT was acquired prior 3rd fraction (one or 2 days post planning CT). Distances from inferior and superior fiducial markers to pubic symphysis plane (perpendicular to vaginal obtulator)were measured. Distance from needle tip of a reference needle (next to the inferior marker) to the pubic symphysis plane was also determined. The difference in fiducial marker distance or needle tip distance between planning CT and CT prior 3rd fraction were measured to assess markers migration and needle displacement. Results: The mean inferior marker displacement was 4.5 mm and ranged 0.9 to 11.3 mm. The mean superior marker displacement was 2.7 mm and ranged 0 to 10.4 mm. There was a good association between inferior and superior marker displacement (r=0.95). Mean averaged inferior and superior marker displacement was 3.3 mm and ranged from 0.1 to 10.9 mm, with a standard deviation of 3.2 mm. The mean needle displacement was 5.6 mm and ranged 0.2 to 15.6 mm. Needle displacements were reduced (p<0.05) after adjusting according to needle-to-fiducials distance. Conclusion: There were small fiducial marker displacements between HDR fractions. Our study suggests a target margin of 9.7 mm to cover interfractional marker displacements (in 95% cases) for pretreatment verification based on radiographic imaging.},
doi = {10.1118/1.4888840},
url = {https://www.osti.gov/biblio/22369746}, journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 41,
place = {United States},
year = {Sun Jun 01 00:00:00 EDT 2014},
month = {Sun Jun 01 00:00:00 EDT 2014}
}