Title: SU-E-T-588: Characterization and Clinical Validation of the Varian Pivotal™ Treatment Solution for Prone Breast Care

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888924· OSTI ID:22369705
; ;  [1]
  1. NY University Medical Center, NY, NY (United States)

Purpose: To report on the clinical validation of the Varian Pivotal™ Treatment Solution for Prone Breast Care: a platform for prone breast radiation therapy. Methods: Patients treated using Breast Conserving Radiation Therapy may benefit from treatment in the prone position with the breast tissue falling freely away from the body. This geometry allows the breast tissue to be treated while avoiding the lung and heart tissue. Eighteen patients simulated and treated using the Varian Medical Systems Pivotal™ Treatment Solution for Prone Breast Care were monitored over the course of treatment for positioning integrity and reproducibility. As this carbon-fiber platform actually replaces a portion of the couch top, indexing is inherent to its design. Patients were positioned on the couch and aligned using fiducial markers and lateral SSD to the breast fiducial point. The daily couch coordinates then serves as indicators for positioning variability with this system. Results: The variations in couch vertical, longitudinal and lateral positions were centered on a mean value of zero with standard deviations of 0.44cm, 0.75cm and 0.79cm respectively. Other factors explored were variations in distance of mid-sternum to table edge and patient rotation into the opening. The median rotation of the chest wall was found to be 11.5 degrees at CT-Simulation with a median distance of 2.5cm from midsternum to support opening. Patient rotation was not associated with either breast size or distance from edge of platform. Conclusion: The Pivotal™ Treatment solution consists of a couch top that replaces the standard top and as such is open from beneath without obstruction. This is a distinction from all other solutions which rely on a platform positioned above and indexed to the treatment couch. We found the reproducibility to be consistent with our historical measures while offering benefits of an integrated solution as stated above. supported by Professional Services Agreeement with Varian Medical Systems.

OSTI ID:
22369705
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English